Aligning or realigning the cross hair is a simple process that any user can perform.
ATTENTION
Make sure that your hands and other objects are free and clear of the path of
the laser beam when testing the crosshair alignment. If the crosshairs are mis-
aligned, the beam may travel to unexpected locations inside the weld chamber.
To set the cross alignment it is necessary to make slight adjustments to two allen screws inside the
chamber (Ref. A and B below). These screws require a 3 mm allen key.
Picture 33: Cross hair realignment
Before making any adjustments, make sure the stereomicroscope is firmly attached in place (Ref. C
in Picture 19 below). Also, make sure that the eyepieces are set to the ‘0’ position before making any
alignment changes.
Picture 34: Fastening of the Stereomicroscope
SCREW A
SCREW B