background image

21

and causing damage if something comes loose while you 
are making adjustments. Place a sheet of white paper inside 
the optical tube directly opposite the focuser. The paper will 
provide a bright “background” when viewing into the focuser. 
When properly set up for collimation, your telescope should 
resemble 

Figure 37.

aligning the Secondary Mirror

To adjust the secondary mirror collimation, you will need a 
Phillips screwdriver and a 2mm hex key, or Allen wrench.
You will need to check, and adjust if necessary, four aspects 
of the secondary mirror’s alignment: 

The secondary mirror’s axial position

1. 

The secondary mirror’s radial position

2. 

The secondary mirror’s rotational position

3. 

The secondary mirror’s tilt

4. 
The first three will probably only need to be checked and (pos-
sibly) adjusted once. Thereafter, it is usually only the second-
ary mirror’s tilt that will need to be adjusted occasionally.

adjusting the Secondary Mirror’s axial Position

With the collimating cap in place, look through the hole in the 
cap at the secondary (diagonal) mirror. Ignore the reflections 
for the time being. The secondary mirror itself should be cen-
tered in the focuser drawtube. If it is off-center along the axis 
of the telescope, i.e., positioned too far toward the front open-
ing or toward the rear of the telescope

as it is in

 Figure 35b,

 

you will have to adjust the mirror’s axial position. 
To do so, use the 2mm hex key to loosen the three small 
alignment set screws in the center hub of the 4-vane spi-
der several turns. Now, grasp the mirror holder (the cylinder 
that is attached to the back of the secondary mirror itself) 
with one hand while turning the center screw with a Phillips 
head screwdriver with your other hand 

(Figure 38).

 Turning 

the screw clockwise will move the secondary mirror toward 
the front opening of the optical tube, while turning the screw 
counter-clockwise will move the secondary mirror toward the 

primary mirror. When the secondary mirror is centered axially 
in the focuser drawtube, rotate the secondary mirror holder 
until the reflection of the primary mirror is as centered in the 
secondary mirror as possible. It may not be perfectly cen-
tered, but that is fine for now. Then, tighten the three small 
alignment set screws equally to secure the secondary mirror 
in that position.

adjusting the Secondary Mirror’s radial Position

Like the axial position, the secondary mirror’s radial position 
was set at the factory and will probably not need any adjust-
ing, or if it does, you’ll typically need to do it only once. 
By “radial position” we mean the position of the secondary 
mirror along the axis perpendicular to the focuser drawtube, 
as shown in 

Figure 39.

 This position is changed by adjust-

ing the two spider vane thumb nuts indicated in the figure. 
Loosen one thumb nut, then tighten the opposite one until 
the secondary mirror is centered radially in the drawtube. Do 
not loosen the thumb nuts too much, to avoid having them 

Figure 37. 

This image shows the XX14i properly set up for 

collimation. Note the white paper placed across from the focuser and 

the horizontal orientation of the optical tube. Ideally, the telescope 

should be pointing at a light-colored wall.

Figure 38. 

To center the secondary mirror axially under the 

focuser, hold the mirror holder in place with one hand while adjusting 

the center screw with a Phillips screwdriver.

Figure 39. 

To center the secondary mirror radially in the focuser 

drawtube, make adjustments to the two knurled spider vane thumb 

nuts that are perpendicular to the focuser.

Spider vane 
thumb nuts

Summary of Contents for 9791

Page 1: ...er Support 800 676 1343 E mail support telescope com Corporate Offices 831 763 7000 89 Hangar Way Watsonville CA 95076 Orion SkyQuest XX14i IntelliScope Truss Tube Dobsonian 9791 Providing Exceptional...

Page 2: ...2 Figure 1 The SkyQuest XX14i IntelliScope Truss Tube Dobsonian...

Page 3: ...t use of the telescope 1 Unpacking The telescope is packed in four shipping boxes as detailed below We recommend keeping the original packaging In the event that the telescope needs to be shipped to a...

Page 4: ...yepiece IntelliScope Object Locator Kit Software 9x50 Finder scope Finder scope bracket with O ring Collimation cap Hex keys x3 Washers Cooling fan mounting screws Upper tube section Lower tube sectio...

Page 5: ...r disk 1 Controller cable coil cable 1 Azimuth encoder cable shorter 1 Altitude encoder cable longer 1 Azimuth encoder spacer washer thin 1 4 outer dia 0 015 thick 4 Encoder connector board washers 3...

Page 6: ...10mm long 15mm dia 1 Insertion tube for rubber washers 3 long 3 Counterweight attachment bolts hex head 1 long black 1 Brass azimuth bushing 1 Azimuth axis bolt hex head 2 75 long 2 Fender washers 1 o...

Page 7: ...ical stop knob Fender washers x2 Altitude tensioning metal washer Vertical stop washers 2 thin 2 thick Altitude tensioning Teflon washer Hex lock nut Counterweight mounting bolts x3 Azimuth axis bolt...

Page 8: ...is hefty 14 Dob has the portability and manageability of a smaller telescope Refer to Figure 7 during base assembly When tightening screws tighten them until firm but be careful not to strip the holes...

Page 9: ...9 N R R M B C D L N O H P J E F A H I Q Q G x12 S x12 T x12 Figure 7 Exploded view of the XX14i base...

Page 10: ...liScope Object Locator kit Insert the modular jack on the azimuth encoder board E 8 into the rectangular cutout in the top baseplate D and align the encoder board so the screw lines up with the pre dr...

Page 11: ...les in each side panel The beveled end of the cylinder should be facing away from the side panel Push a bearing cyl inder screw through the side panel and thread it into the built in nut in the cylind...

Page 12: ...able should plug into the jack on the left side of the encoder connector board Figure 14 Connect one end of the altitude encoder cable the longer 20 of the two flat cables in the IntelliScope kit to t...

Page 13: ...d ing or removing one or more of the supplied thick and or thin washers Adjusting the vertical stop is required when using the IntelliScope Object Locator since the optical tube must point precisely v...

Page 14: ...least three full turns of engage ment on the shafts The mirror cell is now almost ready to be installed onto the lower tube section Before doing so check to make sure that the three mir 4 ror retaini...

Page 15: ...tion instructions For the three counterweights thread them onto the counterweight mounting bolts as far as they will go and tighten them lightly against the metal plate Figure 22 Now the upper and low...

Page 16: ...the knobs and holes line up Figure 26 When tightened the knob will clamp the truss connector against the regis tration flats on the upper truss support ring Figure 27 Repeat this for the other three...

Page 17: ...pres sure Note that the tube will not yet be properly balanced since the eyepiece and finder scope are not in place and the altitude tensioning knob has not been installed Select one of the altitude t...

Page 18: ...et and place it over the body of the finder scope until it seats into the groove in the middle of the finder Unthread the black nylon alignment screws on the bracket until the screw ends are flush wit...

Page 19: ...usly Figure 33 The other eyepiece and 1 25 adapter can be placed in the eyepiece rack until they are needed To install the 10mm Sirius Pl ssl eyepiece 1 25 instead 4 of the 2 DeepView eyepiece keep th...

Page 20: ...4i has a small adhesive ring marking its center This center mark allows you to achieve a very precise collimation you don t have to guess where the exact center of the mirror is NOTE The center ring s...

Page 21: ...head screwdriver with your other hand Figure 38 Turning the screw clockwise will move the secondary mirror toward the front opening of the optical tube while turning the screw counter clockwise will...

Page 22: ...the collimation cap is off center The tilt of the primary mirror is adjusted with three spring loaded collimation knobs on the back end of the optical tube bottom of the primary mirror cell these are...

Page 23: ...e the base to tip because of the reduced leverage In this case it is helpful to place your other hand on the base or the optical tube to help guide the tube Setting the Altitude Tension A nice feature...

Page 24: ...he weight of the eyepiece you can adjust it by tightening or loosening the drawtube ten sioning set screw on the focuser which is located just below the focus lock thumb screw Figure 33 Adjust this se...

Page 25: ...e finder scope If your general aim is accu rate the object should appear somewhere in the field of view Make small adjustments to the telescope s position until the object is centered on the finder s...

Page 26: ...lower tube section Place the dust covers on the tops of the upper and lower tube sections The telescope is now disassembled and ready to be transported Note In step 5 of Assembly of the Optical Tube...

Page 27: ...onomical Observing The SkyQuest XX14i IntelliScope Truss Tube Dobsonian pro vides prodigious capability for observing the many wonders of the heavens from the major planets to deep space nebulas and g...

Page 28: ...Moon are bright enough to see clearly even from light polluted city skies for nebulas galaxies and most star clusters the less ambient light there is to reduce contrast the better When it isn t possi...

Page 29: ...an off by unplugging the battery pack Let Your Eyes Dark Adapt Do not expect to go from a lighted house into the darkness of the outdoors at night and immediately see faint nebulas galaxies and star c...

Page 30: ...subject for observation You can see the disk of the giant planet and watch the ever changing positions of its four largest moons Io Callisto Europa and Ganymede Higher power eyepieces should bring out...

Page 31: ...rcular motion then remove any excess fluid with a fresh lens tissue Oily fingerprints and smudges may be removed using this method Use caution rubbing too hard may scratch the lens For larger lenses c...

Page 32: ...across the mirror Then rinse the mir ror under a stream of lukewarm water Any particles on the surface can be swabbed gently with a series of cotton balls each used just one time Dry the mirror in a...

Page 33: ...ower tube section 52 9 lbs Upper tube section 9 4 lbs Truss pole assemblies 6 9 lbs OTA length assembled 61 OTA length disassembled Lower tube section 21 75 long Upper tube section 8 1 long Truss pole...

Page 34: ...turned postage paid to Orion Warranty Repair 89 Hangar Way Watsonville CA 95076 Proof of purchase such as a copy of the original receipt is required This warranty does not apply if in Orion s judgment...

Reviews: