background image

User manual

UM0972150 R1-0

3

Buttons 

1 2 3 4 5 6 7 8

objects   in

monitor are

closer  than

they appear

C3

C2

C1

Figure 2

Figure 1

1. Service menu

Adjustment of the Spectrum scanner settings is done via the service 

menu when an Orlaco Serial monitor is connected. 

To open the 

service menu, simultaneously press the camera selection button 

(1), the minus button (6) and the plus button (7) (see Figure 1). 

The Monitor service menu (see Figure 2) will appear

. The following 

buttons are used to navigate through the menus:

5 - Option/previous menu:

 Return to the previous menu

6 - Minus:

 Go to the next menu option

7 - Plus:

 Go to the previous menu option

8 - Enter:

 Select or enable the chosen option

2. Spectrum Scanner Menu

There are settings in the Spectrum Scanner that can be adjusted. 

We strongly advise not to change settings before consulting a 

Orlaco service technician.

Open the service menu and go to system settings (see figure 3).

Select the Ext. Device Config (see figure 4). 

A warning is shown in the monitor (see figure 5). Press enter to 

continue.

The Spectrum Scanner Rx (Receiver) menu is shown (See figure 6).

The changes made here are done in the receiver of the Spectrum 

Scanner.

If changes will be made, we advise to make the changes in the 

transmitter first, see chapter 2. 5.

2.1. Language

 (see figure 6).

The OSD menu language is set as English by default. If you would 

like to operate the OSD menu in a different language you can 

change it. The OSD menu is available in English, Dutch, German, 

French, Chech, Hungarian, Italian, Polish, Portuguese, Spanish, 

Turkish, Swedish, Finnish, Danish and Norwegian. 

Figure 3

Figure 4

Figure 5

Figure 6

Reviews: