background image

User manual

UM0972150 R1-0

5

2. 3. 2. Heating setpoint 

(see figure 10)

With the build in heating the spectrum scanner will maintain the 

internal temperature in the set value. Value to be set: OFF; -20°C;

-10°C; 0°C; +10°C; +20°C; +30°C; +40°C. Default setting is +10 °C.

2. 4. OSD Settings

In this menu the settings of the OSD of the Spectrum Scanner can 

be adjusted. (See figure 11 and 12).

2. 4. 1. OSD Menu help

Activation of the pop up help information after 10 seconds.

2. 4. 2. OSD Mode

In this menu the selection of the shown information from the 

transmitter in the monitor.

OFF

B+C

: battery indicator and channel indicator are shown.

C+C

: charge indicator and channel indicator are shown.

CHA:

 only channel indicator is shown.

Default setting is B+C, battery indicator and channel indicator 

are shown.

2. 4. 3. OSD horizontal position

With this the position of the OSD can be horizontally adjusted to 

place is in the middle of the screen.

2. 4. 4. OSD vertical position

Please do not change.

2. 4. 5. Debug mode on/off

This is a function that is used only when an Orlaco service techni-

cian is on site and test the equipment. Standard mode is OFF.

When in OSD mode the channel indicator is activated to be shown 

in the monitor, the shown info gives info about the status of the 

connections:

Capital letter “

C

” means the channel is set to a value.

Small letter “

c

” means the channel is set to auto.

Capital letter “

H

” means data communication is ok.

Small letter “

h

” means that there is no data connection.

cHx

: automatic channel mode, data communication ok.

Chx

: set channel, no data communication.

2. 5. Setup Transmitter

Now the Spectrum Scanner Tx (Transmitter) menu is shown (See 

figure 13).

The changes made here are done in the transmitter of the Spec-

trum Scanner. The menu options are the same as described for the 

Receiver

Figure 11

Figure 12

Figure 13

Reviews: