1. Turn power off from POWER ON/
OFF switch.
2. Simultaneously press the SET
button and the POWER ON switch.
The Lock-out icon will flash.
3. Select and press the button of the
operating features you wish to lock
out. The icon of each feature you
lock out will appear. You may lock
out one or all features.
4. Press the SET button when finished
to set lock outs.
Unlocking the Lock-Outs:
Repeat Steps 1 and 2 on how to set the
Lock Out feature. Select the operating
features you wish to unlock and press
the corresponding unlock and press
the corresponding button. The feature’s
icon will disappear from the LCD. Press
the SET button when finished.
The timer is used to monitor the number
of treatment hours a patient receives.
1. Turn power off from the POWER
ON /OFF switch.
2. Press and hold the PAUSE button
and the POWER ON switch simulta-
neously.
3. The display represents the number
of hours the device has been used
(to the closest hour).
4. To reset the timer to 0 hours for the
next patient, press and hold the
– button until 0 hours is displayed.
The device will now begin counting
usage hours from 0. The timer counts
to a maximum of 999 hours after
which is must be reset.
The ARTROMOT
®
-F can be converted
easily:
1. There are 3 medium finger actuators
(black) and one small finger actuator
(blue). Additional finger actuators in
small, medium or large size may be
ordered separately. See the Ordering
Information chart for details.
2. To replace finger actuators, detach
the black bar (6) containing the finger
actuators by removing both fastening
screws with a Phillips screwdriver
(20).
3. With the bar detached, slide the
finger actuators off the bar and
replace or rearrange them for left
or right hand set up.
4. When changing from right to left
hand, or vice versa, remove the
finger actuators and rearrange them
so that the short finger actuator
(blue) is always in position to attach-
ment to the smallest finger.
– Always unplug the device before
cleaning.
– The ARTROMOT
®
-F can be wiped
clean with disinfectant and thus
fulfills the hygienic requirements for
medical equipment.
– Only wipe the motion device itself
with a moist cloth.
– Protect the motion device from inten-
sive UV light (sunlight).
25
Timer
6. Conversion
Replacing /Changing Finger
Actuators
7. Maintenance
PRECAUTION!
No fluids may be allowed to
enter the casing or the hand-held
programming unit.