DIN EN 13485     ORMED Nr. 018 829-01

© ORMED 

6/04

O R M E D   G m b H   &   C o . K G   •   M e r z h a u s e r   S t r a ß e   112   •   D - 7 91 0 0   F r e i b u r g
Phone +49 7 61 4566-01 • Fax  +49 761 4566-5501 • www.ormed.de • E-Mail: info@ ormed.de
St. Paul/USA, Phone +1 800 4402784, Fax +1 651 4157405 • Prague/CR, Phone +420 2 84094650,
Fax +420 2 84094660 • Vienna/A, Tel +43 1 53 20 83 40, Fax +43 1 53 20 83 431

Summary of Contents for ARTROMOT K3

Page 1: ...Operation Manual ARTROMOT K3...

Page 2: ...for length adjust ment of lower leg 7 Power adapter 8 On Off switch 9 Knurled knob for angle of foot inclination 10 Footplate with patient kit 11 Socket for power adapter 12 Knurled knob for rotating...

Page 3: ...7 4 1 Connecting the device 7 4 2 Adjusting the femur length 7 4 3 Adjusting the supports 7 5 Setting the treatment values 8 5 1 Programming the ARTROMOT K3 8 5 1 1 Programming the treatment values 8...

Page 4: ...oma resorbtion the improvement of lymph and blood circulation the prevention of thrombosis and embolism 4 1 How to use ARTROMOT K3 1 3 Indications 1 1 Application 1 2 Objectives of therapy The ARTROMO...

Page 5: ...nd store any treatment values 5 Note Fold out page 1 1 Knob for femur length adjustment 2 Base 3 Coiled cord 4 Controller cable 5 Hand held programming unit 6 Knob for length adjustment of lower leg 7...

Page 6: ...necessary Make sure that the characteristic values of your power supply correspond to the voltage and frequency data indicated on the ID plate Only connect the ARTROMOT K3 to correctly installed safet...

Page 7: ...move the foot support horizontally and adjust precisely to the patients lower leg length figure 3 Positioning of foot dorsi plantar flexion Loosen the two knobs 9 and adjust the foot plate at a comfor...

Page 8: ...n or flexion and can be set separately for extension and flexion Possible values for pauses 0 30 seconds 5 1 2 Information about treatment values CAUTION The programmed value and the actual angle meas...

Page 9: ...ons press the STOP key Press the START key the device checks programmed values Adjusting the force reverse on load Minimum setting for reverse on load 25 kp Maximum setting for reverse on load 45 kp S...

Page 10: ...y 20 to 85 Air pressure 700hPa to 1060 hPa Operational Conditions Surrounding temperature 50O F to 104O F Relative humidity 30 to 75 Air pressure 700hPa to 1060 hPa Electrical rating 115 230V 50 60Hz...

Page 11: ...com Technical hotline Do you have any technical questions Do you need technical service Tel 49 180 51 ormed de 49 180 51 676 333 Fax 49 180 53 ormed de 49 180 53 676 333 CAUTION Carry out regular chec...

Page 12: ...g units Type Knee Hip Name ARTROMOT K3 meets all requirements of following EC directives EN 60 601 1 1990 Electrical Medical Devices Part 1 Basic Rules for Safety EN 60 601 2 1993 Electrical Medical D...

Page 13: ...17 Figures Fig 2 Fig 3 Fig 4 Fig 5 Fig 6 Fig 7 Fig 8...

Page 14: ...h a u s e r S t ra e 112 D 7 910 0 Fre i b u r g Phone 49 761 4566 01 Fax 49 761 4566 5501 www ormed de E Mail info ormed de St Paul USA Phone 1 800 4402784 Fax 1 651 4157405 Prague CR Phone 420 2 84...