3. Safety instructions

- The ARTROMOT

®

-K3 may only be 

operated by authorized persons.

- Make sure that the patient is supported

in an anatomically correct way. Check
the following settings/ positioning:

1. Femur length

2. knee joint axis

3. calf length and leg rotation setting

4. patient kits

- In case of patients who are adipose, 

particular large or very small, you 
should pay attention to the following:

- Avoid abrasion and pressure

- If necessary support the leg in a 

slightly abductive position

- The maximum continuous load on the 

leg support element is 30 kg

- Movement must always be free of 

pain and irritation

- The patient must be fully conscious

during instruction and when using the 
splint.

- The doctor or therapist must decide 

on a case-to-case basis wether the 
device can be used with the patient

- The hand-held programming unit 

should be explained to the patient 
and must be located within the 

patient´s reach, so that the therapy 
can be interrupted if necessary.

- Make sure that the characteristic 

values of your power supply
correspond to the voltage and 
frequency data indicated on the ID 
plate.

- Only connect the ARTROMOT

®

-K3

to correctly installed safety sockets.

Repair and maintenance work may 

only be carried out by authorized 
persons, as otherwise all warranty 
services and liabilities shall be void.

- Perform regular checks on all com-

ponents for possible damage or loose
connections

- Damaged or worn parts should be 

replaced immediately with original 
spare parts by an authorized specialist.

- Before cleaning and repair disconnect

the device from the main socket.

- When carrying out any work on the 

device, never allow liquids to get 
inside the housing or the hand-held 
programming unit.

- Only use the AC-AC adapter supplied 

with the unit.

To disconnect the device from mains,
unplug the AC-AC adapter from the wall
socket.

CAUTION!

these instructions must be read 
before start- up!

CAUTION!

Before treatment begins, a test run
involving several movement cycles 
should be carried out first without 
and then with the patient.

CAUTION!

The ARTROMOT

®

-K3 may only be 

operated with the attached power 
supply NTEV20

Summary of Contents for ARTROMOT K3

Page 1: ...Operation Manual ARTROMOT K3...

Page 2: ...for length adjust ment of lower leg 7 Power adapter 8 On Off switch 9 Knurled knob for angle of foot inclination 10 Footplate with patient kit 11 Socket for power adapter 12 Knurled knob for rotating...

Page 3: ...7 4 1 Connecting the device 7 4 2 Adjusting the femur length 7 4 3 Adjusting the supports 7 5 Setting the treatment values 8 5 1 Programming the ARTROMOT K3 8 5 1 1 Programming the treatment values 8...

Page 4: ...oma resorbtion the improvement of lymph and blood circulation the prevention of thrombosis and embolism 4 1 How to use ARTROMOT K3 1 3 Indications 1 1 Application 1 2 Objectives of therapy The ARTROMO...

Page 5: ...nd store any treatment values 5 Note Fold out page 1 1 Knob for femur length adjustment 2 Base 3 Coiled cord 4 Controller cable 5 Hand held programming unit 6 Knob for length adjustment of lower leg 7...

Page 6: ...necessary Make sure that the characteristic values of your power supply correspond to the voltage and frequency data indicated on the ID plate Only connect the ARTROMOT K3 to correctly installed safet...

Page 7: ...move the foot support horizontally and adjust precisely to the patients lower leg length figure 3 Positioning of foot dorsi plantar flexion Loosen the two knobs 9 and adjust the foot plate at a comfor...

Page 8: ...n or flexion and can be set separately for extension and flexion Possible values for pauses 0 30 seconds 5 1 2 Information about treatment values CAUTION The programmed value and the actual angle meas...

Page 9: ...ons press the STOP key Press the START key the device checks programmed values Adjusting the force reverse on load Minimum setting for reverse on load 25 kp Maximum setting for reverse on load 45 kp S...

Page 10: ...y 20 to 85 Air pressure 700hPa to 1060 hPa Operational Conditions Surrounding temperature 50O F to 104O F Relative humidity 30 to 75 Air pressure 700hPa to 1060 hPa Electrical rating 115 230V 50 60Hz...

Page 11: ...com Technical hotline Do you have any technical questions Do you need technical service Tel 49 180 51 ormed de 49 180 51 676 333 Fax 49 180 53 ormed de 49 180 53 676 333 CAUTION Carry out regular chec...

Page 12: ...g units Type Knee Hip Name ARTROMOT K3 meets all requirements of following EC directives EN 60 601 1 1990 Electrical Medical Devices Part 1 Basic Rules for Safety EN 60 601 2 1993 Electrical Medical D...

Page 13: ...17 Figures Fig 2 Fig 3 Fig 4 Fig 5 Fig 6 Fig 7 Fig 8...

Page 14: ...h a u s e r S t ra e 112 D 7 910 0 Fre i b u r g Phone 49 761 4566 01 Fax 49 761 4566 5501 www ormed de E Mail info ormed de St Paul USA Phone 1 800 4402784 Fax 1 651 4157405 Prague CR Phone 420 2 84...

Reviews: