11. Checklist of safety and function test
ARTROMOT®- K1 Comfort and Standard version
Safety test
Protective earth conductor resistance
Ground leakkage current EN 60601 / IEC 601/ VDE 0751
Or
Ground leakkage current as in UL 2601
Function test
1.
Measured value
Date/ Signature
0,1 Ohm
Ohm
500 µA
µA
300 µA
µA
OK
Error
≤
≤
≤
Switch on the ARTROMOT® -K1. Press the two outer buttons simultaneously.
Display: Software version VX.X XX.XX.XX (X = optional) Keep on pressing.
Display: ARTROMOT K1
“Product version”
2. The maximum range of motion for Extension/ Flexion is -10 to 120 degrees.
Check the angle in position 0 degrees. Tol/- 5 degrees.
Check the angle in position 60 degrees. Tol/- 5 degrees.
Check the angle in position 100 degrees. Tol/- 5 degrees.
3. Check the emergency-off function.
Start the ARTROMOT® K1 in any mode.
Press any key, the ARTROMOT® K1 will stop immediately.
Check this for all keys.
4. Set up the special function “new patient“
press to activate it.
Press
START
. The ARTROMOT® -K1 will move automatically to 30 degrees
and stop there.
5. Check the set values of “new patient”
Extension
Flexion
Speed
Reverse
Therapy Duration
->
->
->
->
->
Display: 25°
Display: 35°
Display: 50 %
Display: 25
Display: 00:00
6. Start the ARTROMOT® -K1 in the motion range between –10 to 120 degrees.
Set the speed
to 100%.
Both extreme points should be reached within 45 – 65 seconds.
7. All special funtion are deactivated. Select a therapy time of 3 minutes.
At the end of the 3 minutes the ARTROMOT®-K1 will stop automatically in
the middle position.
15