- 23 -
6. 8 HOW TO USE CHECKER SWITCH
1. How to detach the Auto-Winding Module.
Detach the auto-winding module from Osada Apit 11 body
by pressing its bottom.
2. How to use the Checker Switch.
1) Confirm the checker switch behind the Osada Apit 11 body is on
the
“MEASURE” position.
2) Turn on the power by depressing the power switch.
CAUTION
The checker functions with the auto-adjustment mode and the manual-
adjustment mode.
3) Slide the checker switch to the ADJ. position. Press the ADJ. switch
set on the body and confirm if the meter needle on the display
panel is pointing within ±2 scale range from the ADJ. position ,
as shown in the right figure.
4) Next, move the checker switch to the W.L. position and confirm the
meter needle on the display panel if it points within ±2 scale range
from the W.L. position on the display panel, as shown in the right
figure.
CAUTION
Do not skip 3) above, always take both steps of 3) and 4) consecutively.
If you go to the step 4) directly, the checker does not function properly.
6 OPERATIONS
Its Bottom
Position of the switch
Power Switch