- 25 -
7. 2 STERILIZATION
Steam Autoclave
Items which can be steam-autoclaved are to be referred to the list of
“STERILIZATION”
on Page
25.
TEMPERATURE
DURATION
121°C
20 min. and more
132°C
5 min. and more
135°C
3 min. and more
Regarding the possibility to re-sterilize the lip clip and file holder cord, they can be
sterm-autoclaved up to two hundred fifty (250) times.
CAUTION
Always put in a sterilization bag when Steam Autoclave is given. Do not utilize
the drying process
7. 3 METHOD FOR CLEANING, DISINFECTION AND STERILIZTION (REFERENCE CHART)
METHOD
ITEM
CLEANING
DISINFECTION
STERILIZATION
WIPING WITH NEUTRAL
DETERGENT OR WATER
WIPING WITH
ETHANOL
STEAM
AUTOCLAVE
Osada Apit 11 Body
O
X
X
Auto-Winding Module
O
X
X
Probe Cord*
O
O
X
File Holder Cord
O
O
O
Lip Clip
O
O
O
7 CLEANING, DISINFECTION AND STERILIZATION
Lip Clip
File Holder Cord