- 28 -
11. 1 GUIDE FOR REPAIRS
The meter needle on the display panel swings so soon as the power is turned on.
1) The meter needle on the display panel swings out.
The probe cord of the manual-winding module can be changed by you, while
the probe cord of the auto-winding module cannot be changed by you.
2) The meter needle on the display panel stops in the middle and does not move at all.
YES
NO
YES
NO
YES
YES
NO
The checker switch behind Osada
Apit 11 body in on the ADJ. or W.L.
position.
Return the checker switch behind
Osada
Apit
11
body
to
the
MEASURE
” position.
The auto-winding module (or man-
ual-winding module) is not being
attached properly.
Attach the auto-winding module (or
manual-winding module) to the
Osada Apit 11 body securely.
When detaching the auto-winding
module (or manual-winding module)
from Osada Apit 11 body, the meter
needle returns to the left completely.
Change the auto-winding module
into the new one, or change the
probe cord of the manual-winding
module into the new one.
NO
Stop using the device and call for repair.
The set icon, , appears on the
display panel and the meter needle
blinks.
Wait for three(3) seconds. The set
icon, , disappears and the
meter needle stops blinking.
NO
The lip clip and metallic clipping end
of the file holder cord contact.
Separate the lip clip from the
metallic clipping end of the file
holder cord.
YES
NO
The auto-winding module (or man-
ual-winding module) is not being
attached to the Osada Apit 11 body
properly.
Attach the auto-winding module (or
manual-winding module) to the
Osada Apit 11 body securely.
YES
NO
When detaching the auto-winding
module (or manual-winding module)
from Osada Apit 11 body, the meter
needle returns to the left completely.
Change the auto-winding module
into the new one, or change the
probe cord of the manual-winding
module into the new one.
YES
Stop using the device and call for repair.
CAUTION