11. 1 GUIDE FOR REPAIRS
3) The meter needle on the display panel does not swing or swings less.
Power is not supplied.
Turn on the power switch.
YES
NO
The auto-winding module (or man-
ual-winding module) is not being
attached properly.
Attach the auto-winding module (or
manual-winding module) to the
Osada Apit 11 body securely.
YES
NO
The meter needle does not swing
when contacting the lip clip and the
metallic clipping end of the file holder.
Change the auto-winding module
into the new one, or change the
probe cord of the manual-winding
module into the new one.
YES
NO
Stop using the device and call for repair.
NO
The meter needle does not point within
±2 scale range when confirming the
positions of the
“ADJ.” and “W.L.” with
the checker switch.
Stop using the device and call for
repair.
YES
NO
The metallic clipping end of the file
holder cord is dirty or corroded.
Clean the metallic clipping end of
the file holder cord, or change the
file holder cord with the new one.
YES
11 REFERENCE