- 8 -
ITEM
SPECIFICATION
REMARKS
1. Product
Osada Apit 11
2. Model (Model Number)
EM-S11
3. Classification according to the type
of protection against electric shock
Internally powered equipment
EN60601-1
4. Classification according to the degree
of protection against electric shock
Type BF applied part
A
pplied part
:
File Holder Cord
、
Lip Clip
EN60601-1
5. Classification according to the degree
of protection against water perme-
ability
IPX0
6. Operating Mode
-Continuous operation
-The device meant to be used as transient
(the continuous duration is limited to less
than 60 min.)
7. Precautions against flammable
atmospheircs
This device is not suitable to be used in the
air/flammable anesthetic gas, oxygen,
nitrous oxide, or flammable anesthetic gas.
8. Power
Source
①
Power Voltage
DC 6 V
Applicable Battery
4 pcs. of AA-size alkaline dry batteries
Power Input
15mA
9. Power
Source
②
Power Voltage
DC 4.8 V
Applicable Battery
4 pcs. of AA-size rechargeable nickel
hydride batteries
Power Input
20mA
10. Body Dimensions
100 (W)mm x 79 (D)mm x 101 (H)mm
11. Body Weight
420 grams (excluding batteries)
12. Environmental Conditions (IN USE)
Ambient Temperature
… 10 ~ 40ºC
Relative Humidity
……… 30 ~ 75%
Atmospheric Pressure
…700 ~ 1,060hPa
Less than 2,000m altitude
13. Software Version
Ver 1.00
14. User (Dentist)
-The dentist who has normal vision and
normal audibility (the dentist who has the
slight impairment in vision or audibility or
the vision corrected to more than log
MAR0.2 (6/10 or 20/32) or audibility
corrected to be able to hear 30dB is
accepted.).
-No particular education or training
required for operating the device.
-Should read and understand English and
the Arabic figures.
4 SPECIFICATIONS