Servicing instructions
Always unplug the mains cable when undertaking repair and maintenance!
If the mains cable is damaged it must be replaced. New ones can be bought from the
supplier.
Mounting the refrigeration unit
1.
Unplug the power cable from the wall socket.
2.
Disconnect the sensor cable from the mounting bracket on the small bottle.
3.
Remove the hood screw and remove the hood.
4.
Remove the four screws holding the unit in place.
5.
Lift the unit straight up. Take care to remove the sensor cable as well; there is a risk that it can
be pulled apart.
6.
Mount the new unit in the reverse order.
Installing the medication and pharmacy cabinet
If the refrigerator needs to be moved or repositioned it is recommended that the cabinet be moved
standing upright. The safest way to move or reposition the cabinet is described below. Use a sack
truck.
1.
Unplug the power cable from the wall socket.
2.
Remove the medication and shelves from the refrigerator.
3.
Remove the defrost water container and glycerol bottle.
4.
Remove the anti-tip protection from the wall.
5.
Move the refrigerator to the desired location. If the cabinet is to be transported in a vehicle the
cabinet should be packaged.
6.
If there are adjustable feet these should be mounted.
7.
Secure the cabinet with the anti-tip protection.
8.
Insert the shelves in the desired positions.
9.
Install the defrost water container and glycerol bottle.
10.
Plug the power cable into the wall socket.
If it is necessary (despite the recommendation to keep it upright) to transport the cabinet horizontally,
take the following precautions.
1.
Horizontal transport must only be undertaken internally and then only for short distances!
2.
Remove the medication and shelves from the refrigerator.
3.
Remove the defrost water container and glycerol bottle.
4.
The refrigerator must be transported on its left-hand side.
5.
The cabinet must stand upright for at least two hours before it is switched on.
6.
If the cabinet has been exposed to low temperatures during transport it must stand upright for
at least 6 hours before it is switched on.