Installation and operating manual for WATERFRIEND -Chlorine MRD-3 Page 37
Calibration of the temperature sensor
With this feature, the temperature sensor in the flow cell can be calibrated.
Procedure:
Press the
key
1. Select Service functions
2. Press the key
in the key bar
3. Press the key
in the key bar once again
4. Select Temperature correktion.
5. By pressing the arrow keys
or
, or with the keyboard on
the right of the display, the actual temperature value measured
with a thermometer is entered. For confirmation, press the
button!
Additional settings via the LAN interface
The embedded Web server provides additional settings that affect Internet communication.
To change these settings, the web server must be opened in the web browser on your computer.
After that, you need to log in by clicking on the key icon
and entering the LAN PIN or the service PIN:
Key icon for login
After login, you can go to the page "Menu -> Service functions ->
Network Settings" and, enter a name for the device. This name appears
in future network accesses in the title bar of the web browser, and also
appears in the subject line of any emails sent.
You can also specify two recipients for error message emails.