7.5
Special Grounding Note:
The unit must have an earth driven ground rod within five (5) feet of the hospitable power
receptacle. A grounding cable of a minimum #3/0 AWG must be connected between the grounding
rod and the grounding pin of the hospital power receptacle. If required by local codes, another
cable, to be kept as short as possible, may also be connected between the ground stud on the
Incoming Power Distribution Panel and an earth driven ground rod.
Figure 37: Ground Connection