9.7 Humidity
Controller
All settings for the humidity system are preset at the factory. Under no
circumstances should factory presets be altered.
The humidity controller is located in Equipment Room as part of the Chiller Control Panel. The
relative humidity setting for the mobile unit is 50%. The humidifier must not be altered from its
factory setting.
9.8 Humidity
Settings
All settings for the humidity system are preset at the factory. Under no
circumstances should factory presets be altered.
The humidity low set point is 30% RH (relative humidity).
The humidity high set point is 60% RH (relative humidity).
9.9 Electrical
Connections
Electrical connections at the humidifier are located on a terminal rail behind the cover of the
humidifier.
The distribution panel supplies the required 480V AC power via a 15 amp, 3-phase breaker.
9.10 Instructions
The HVAC system along with the humidifier is set to the required settings per the medical
equipment manufacturers’ specifications before leaving the factory. Under no circumstances should
the settings be altered from their factory specifications.
For additional information, refer to the Oshkosh Specialty Vehicles VOL II Vendor Information
binder for the product manual.