37
4 OFS System Configuration Setup
4.1 Checking Communications
At this point you should have serial communications
established between the computer and the OFS as
described in Section 3.14
.1. ”HyperTerminal Setup”.
Check the status bar at the bottom of the
HyperTerminal window. You will see connection
duration, bit rate configuration and other indicators.
The OFS responds to uppercase ASCII character
inputs as described elsewhere in this manual, but for
now we will begin by entering “V”. The system will
respond with firmware version and install date. This
shows that the system is indeed on
– line and
responding.
4.2 OFS 2000C Setup
Command Conventions
OFS responds only to Uppercase Characters.
Pressing the <Enter> key at any time will exit the
Setup function.
Change commands will be acknowledged in two
steps
“Erasing Save Buffer – Done” shows the
change command has been accepted
“Done” shows the change command has
been effected.
Setup
Setup is initiated with an “S” character.
A list of setup parameters “0” through “9” will appear.
4.2.1 Units of Measurement
You may choose
Note: Choosing Metric or English Units here will
automatically switch all system measurements to the
same matching UOM.
Enter the number for the preferred UOM. The
following message will appear to show the command
has been accepted.
Erasing Save Buffer
Done
Done
(You may enter “S” again to display the list and verify
that the parameter has been set.)
Figure 4.1 HyperTerminal Open
Figure 4.2 "V" Poll Output
Figure 4.3 Units of Measure
0: m/s
Meters per Second (Default)
1: kph
Kilometers per Hour
2: mph
Miles per Hour
3: fps
Feet per Second
4: fpm
Feet per Minute