Pressure
P
Fixed Field - Pressure
62
,
Comma delimiter
63-66
pppp
Pressure in hPa (1013 hPa = 1 Atm)
67
,
Comma delimiter
68
Heater Rod Current
K
Fixed Field
– Heater Current
69
,
Comma delimiter
70-73
kkkkk
Heater Current in Amp
* If some of the Aux inputs are disabled, the corresponding data fields are dashed out so that it does not
cause fault condition.
5.2.2 Description of Status Indicator Codes (Bytes 29-32)
Byte Unit of Measure Description
Byte
Averaging Time
Description (sec)
29
0
m/s
30
0
10
1
kph
1
30
2
mph
2
60
3
fps
3
120
4
fpm
4
300
5
600
6
3*
Byte Operation Mode
Description
Byte
Full Scale Range
Description
31
0
Normal Operation
32
0
0-40 m/s
1
A/B out of range
1
0-20 m/s
2
Velocity out of the
range defined by
Byte 32
2
0-10 m/s
3
N/A
3
0-5 m/s
4
Calibration Mode
5
Heater >25% current
6
RTD Fault
7
Pressure Sense Fault
8
OFS restart
9
Clean Windows