10 Maintenance
10.1 Safety
This equipment does not contain any operator replaceable parts. Only trained personnel are allowed to service
this equipment.
This equipment contains hazardous voltages. Disconnect AC power before servicing. Read the entire User's
Guide before proceeding with installation or maintenance!
10.2 Preventive Maintenance
The following preventative maintenance (PM) actions are recommended:
Monthly
Record the information displayed on the controller monthly in a dedicated PM log book for each meter. (This
helps establish the rate of deposition of dirt on the windows and therefore your window cleaning PM work. It
also helps when you have to diagnose a problem which has appeared and you want to compare with the way
it used to look)
Biannually
(or when cleaning windows)
Installation condition check
– Inspect the TX Head, RX Head and Control Box for proper installation.
Verify that all mounting bolts are installed and secure. Verify that the TX/RX Head mounting gaskets are
installed, in good condition, and not leaking air. Check the power and signal cables to verify that they are
tight and not frayed.
See that the purge air (if applied) is flowing properly.
Check the aim
Visual indicator check
– Verify that the LED indicators are illuminated per Section 9.2 “Visual Indicators
Check
”.
Window fouling check
– Using the Control Box digital display, verify that the A & B voltages are > 0.5
volts or have not decreased significantly from the last PM check. If less than 0.5 volt, follow the steps in
Section 10.2.1
“Window Cleaning Procedure” to clean the windows.