OCT/SLO SERVICE MANUAL Version V.4
7
ELECTRICAL
Imaging System only:
Single phase:
100V ~ systems: (
±
10%), 3.5A
115V ~ systems: (
±
10%), 2.6A
230V ~ systems: (
±
10%), 1.3A
Total Power requirements and Power consumption for:
Imaging system, PC computer, LCD monitor and Motorized Table:
100V ~ : (
±
10%), 12.0A 1KVA
115V ~ : (
±
10%), 10.6A 1KVA
230V ~ : (
±
10%), 5.3A 1KVA
Main fuses: for 100V ~ & 115V 2 each 8.0A. for 230V ~ 5.3A
CONTROL UNIT
(Specifications current at time of writing)
CPU:
2.6GHz Pentium 4, 1GB DDR RAM
Monitor:
19” Color LCD Monitor
Control Input Devices:
Keyboard, Mouse, Joystick
Storage:
160GB Hard Disk
ENVIRONMENTAL CONDITIONS
Transport and Storage
Temperature:
–10 to +60 deg. C
Relative Humidity:
30% to 70%, excluding condensation
Atmospheric Pressure: 500 hPa to 1060 hPa
Operation
Temperature:
+17 to +27 deg. C
Relative Humidity:
30% to 70%, excluding condensation
Atmospheric Pressure: 700 hPa to 1060 hPa
DIMENSIONS:
Scanner:
59x67x55cm
41kg
Power Supply:
22x48x20cm
16kg
Note: System specification may change from time to time by the
manufactuere to to improvamnet in components performance or
other technical reasons.