86
87
About Start-up Handling Options Tinnitus Warnings
More info
215273US / 2019.11.21
215273US / 2019.11.21
Technical Data
0 dB SPL ref. 20 mPa
Oticon
Opn 1, Opn S 1,
Opn Play 1
Oticon
Opn 2, Opn S 2,
Siya 1, Ruby 1
Oticon
Opn 3, Opn S 3, Opn
Play 2, Siya 2, Ruby 2
Peak OSPL90
131 (1291) dB SPL 131 (129*) dB SPL 131 (129*) dB SPL
HF Average OSPL90
126 (118
1
) dB SPL 126 (118*) dB SPL 126 (118*) dB SPL
Peak Full-on Gain
66 (66
1
) dB
66 (66*) dB
66 (66*) dB
HF Average Full-on Gain
61 (54
1)
dB
61 (54*) dB
61 (54*) dB
Reference Test Gain
50 dB
50 dB
50 dB
Frequency Range
120-7000 Hz
120-7000 Hz
120-7000 Hz
Total Harmonic Distortion 500 Hz
3 %
3 %
3 %
Total Harmonic Distortion 800 Hz
<2%
<2%
<2%
Total Harmonic Distortion 1600 Hz
<2 %
<2 %
<2 %
Battery Consumption
1.9 mA
1.9 mA
1.9 mA
Equivalent Input Noise Level
(omni/dir)
14/27 dB SPL
14/27 dB SPL
14/27 dB SPL
HF Average SPLITS (left/right ear)
109/109 dB SPL 109/109 dB SPL
109/109 dB SPL
Attack Time
<5 ms
<5 ms
<5 ms
Release Time
<50 ms
<50 ms
<50 ms
2CC Coupler
measured
according to
American
National
Standard
ANSI S3.22-2014
and ANSI
S3.55-2014/
Part 5
Supply voltage:
Battery Zinc
Air 1.4 Volt
Oticon
Opn 1, Opn S 1 & Opn Play 1
Oticon
Opn 2, Opn 3, Opn S 2, Opn S 3, Opn Play 2,
Siya 1, Siya 2, Ruby 1, Ruby 2
OSPL90 – Output
Sound Pressure Level
Input: 90 dB SPL.
Technical setting: A0
Full-on Gain
Input: 50 dB SPL.
Technical setting: A0
BTE PP
100
110
120
130
dB SPL
100
200 Hz
500
1000
2000 Hz
5000
10000
Opn1BTE13PP_OSPL90_2cc − 02−05−2017 − taje − 09:02:09
100
110
120
130
dB SPL
100
200 Hz
500
1000
2000 Hz
5000
10000
Opn2+3BTE13PP_OSPL90_2cc − 02−05−2017 − taje − 13:28:59
40
50
60
70
dB
100
200 Hz
500
1000
2000 Hz
5000
10000
Opn1BTE13PP_FOG_2cc − 02−05−2017 − taje − 09:03:26
40
50
60
70
dB
100
200 Hz
500
1000
2000 Hz
5000
10000
Opn2+3BTE13PP_FOG_2cc − 02−05−2017 − taje − 13:31:47
Sound Hook
Thin tube
(size 1/1.3)
* For instruments
fitted with thin
tube