30
OTICOn InTIGA FITTInG GUIDE
31
Intiga is shipped ready for the selected Receiver and the
Programming Top Shell to be positioned. In the nOAHlink
Properties Advanced -- Advanced Settings menu, insure
that the “HI Power State” is On.
Do nOT insert a battery into the Intiga when using the #3
programming cable. The power to Intiga will be provided by
the programming cable.
1.
Attach the Receiver by aligning the pins in the base shell
with the Receiver.
2.
Attach the Programming Top Shell by inserting the front
hooks into the base shell and lower the other end until it
snaps into the base.
Programming Oticon Intiga the first time
3.
Attach the #3 programming cable to the top shell.
Thread the cable through the ‘tail’.
Use of the Programming Top Shell is
vital for accurate programming of
Intiga. The Programming Top Shell
ensures accurate operation of the
Feedback Manager and In-Situ
Audiometry. Omitting the Program-
ming Top Shell will affect the acous-
tics and sound quality of Intiga when
using the Fitting Assistant. Lastly,
the use of the Programming Top Shell
ensures a reliable and stable connec-
tion during the fitting process.
Important Note