OTICOn InTIGA FITTInG GUIDE
41
Controls
Adjusting the fitting controls
By clicking the
<>
each or all of the three regions (LF, MF
and HF) can be expanded to display all the underlying
frequency channels.
Click
to display the MPO controls for each region or
channel.
Adjust the gain controls in 1 dB steps in the selected fields.
MPO is displayed in dBHL while gain is displayed in dB (IG).
Reserve Gain
The VC Reserve Gain displays the volume range between
the preferred level and the instrument’s possible maximum
gain in the range of 0-10 dB, at 2.5 dB steps. To modify the
Reserve Gain, adjust the instrument’s gain or change the
feedback limit.
The volume control ranges 30 dB. From the preferred level,
gain can be increased 10 dB and decreased 20 dB.
Compression Ratios
Click on the ‘CR’button between the right and left graphs to
read the compression ratio.
Program Manager
In the Program Manager you can activate, configure, delete,
and change the order of the programs. For general programs
you can choose between the VAC, nAL nL1, nAL nL2 and
DSL v5.0 fitting rationales.
If you are fitting ConnectLine to your Oticon Intiga instru-
ments, adjustments to ConnectLine programs are made
in the End Fitting step.