Volume Control
allows on-screen adjustment of the instrument's VC during the fitting. Manual
operation overrides the on-screen setting of the VC and vice versa.
VC range
sets the available range between position 1 and 4 of the volume control. Choices are
Full (30 dB), Half (15 dB) and Disabled (no range). The Full range is the default setting.
Fitting tips:
1)
Before reducing the range, make
sure that appropriate gain is fitted in
the Full range, and that the VC is set
to preferred level.
2)
When changing VC Range, consider
the listening environments the client
is exposed to during the day. If the
loudness of the sound level changes
often and varies widely, then “Half”
range might not be a good solution,
because the client might not be able
to sufficiently decrease the volume
if the enviroment becomes uncomfort-
ably loud.
3)
Clients who have difficulty with
adjusting inside the “Full” range,
and who are not exposed to varied
listening situations, might benefit
from “Half” range or “Disabled”
setting.
11
Step 2d: Check Programmable
Volume Control