INSTALLATION AND STARTUP
24
OI-6000-X-X-X-X-O OPS_GUIDE_REV 3.1
2.5
NORMAL OPERATING MODE
During normal operating mode, the OI-6000 continuously samples the air and updates the measured concentration
of the target gas on the display screen. The display, when in normal operation, appears as shown below.
1
Measured gas concentration (reading)
2
Gas concentration unit of measure
3
Sensor element gas type
4
Readings above full scale show as MAX and the 4-20 output is set to 25 mA
In the event of a device failure, the unit will alternate between the normal operating screen and a fault screen on the
display, in 5 second intervals, until the fault has been cleared, or is corrected. The fault code, located in the bottom-
left corner of the display, appears on both screens. The unit continuously registers that the system is in fault, so that
even with at-a-glance instrument checks in the field, it can be seen. When the fault is corrected, the unit will return
to normal operating mode.
For a list of the fault codes and warning symbols of the OI-6000, and their associated meaning, refer to the Product
Troubleshooting section of this manual.
Both system menus are accessible from the normal operating mode. To access the product settings and
configuration menu, press and hold the MENU button, for approximately 6 seconds, until the menu is activated and
open on the display screen. To access the operation settings menu from the normal operating screen, press the
MENU button once and the menu will open and show on the display.
NOTICE
After 5 minutes of no interaction with the device, the unit will automatically return to normal operating mode.