OPERATION SETTINGS
OI-6000K-X-X-X-NXP-X OPS_GUIDE_REV 1.0
40
4.2.3
CALIBRATING THE SENSOR (AUTO CAL)
You should
ONLY
perform the calibration of the sensor after the null process has been completed.
For best results, use 50% of the sensor scale of your target gas in an air balance with a flow rate
of 0.25 to 0.5 LPM.
1.
Press the
ADD
button to select “Yes” to begin the calibration process and to advance to the
calibration confirmation screen. If you do not wish to calibrate the sensor, press the
SUB
button to select “No” to advance to the sensor radio address setting screen.
2.
Press the
ADD
button to select “Yes” to confirm that you want to calibrate the senso
r and to
continue to the concentration setting screen. If you do not wish to continue to calibrate the
sensor, press the
SUB
button to select “No” to advance to the sensor rad
io address setting
screen.
3.
Use the
ADD
and
SUB
buttons to adjust the concentration to the calibration gas being used.
Press the
MENU
button to save the gas concentration setting and to advance to the
calibration start screen.
4.
Affix a Calibration Adapter Kit (sold separately) to the sensor housing of the device.
Summary of Contents for OI-6000K NXP Series
Page 1: ......
Page 2: ......
Page 67: ...APPENDICES OI 6000K X X X NXP X OPS_GUIDE_REV 1 0 62 Notes ...