PRODUCT MAINTENANCE
OI-6000K-X-X-X-NXP-X OPS_GUIDE_REV 1.0
46
5.2
SENSOR REPLACEMENT
The sensor elements used in the OI-6000K detects gas in either % or PPM concentrations, this element must
be fully functional in order for the system to operate correctly. Otis recommends replacing the sensing
element whenever a slow response to gas is observed during the normal calibration process. After replacing
the sensing element the device
MUST
then be nulled and calibrated for proper operation of the device.
CAUTION
The internal components can be static sensitive. Use caution when opening the enclosure and
handling internal components.
DO NOT
use any metal objects or tools to remove the sensing element from the sensor adapter
board.
1.
Press and hold
the
SUB
button for approximately 6 seconds, until “OFF” shows on the display
screen.
2.
Unscrew and remove the sensor housing cap from the sensor housing base. Set aside.
3.
Using the thumb and forefinger, gently unplug the sensing element from the sensor housing
board.