Otometrics
107
App. 3 Technical Specifications
Notes on EMC (Electromagnetic Compatibility)
Guidance and manufacturer's declaration - electromagnetic emissions for all equipment and systems
AccuScreen is intended for use in the electromagnetic environment specified below. The user of AccuScreen should ensure that it
is used in such an environment.
Emissions
test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
AccuScreen uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B
AccuScreen is suitable for use in all environments, including domestic environments and
those directly connected to public low-voltage power supply networks that supply build-
ings used for domestic purposes
Guidance and manufacturer's declaration - electromagnetic immunity for all equipment and systems
AccuScreen is intended for use in the electromagnetic environment specified below. The user of AccuScreen should ensure that it
is used in such an environment.
Immunity test
IEC 60601
test level
Compliance level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
+/- 6 kV contact
+/- 8 kV air
+/- 6 kV contact
+/- 8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humid-
ity should be at least 30 %.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels char-
acteristic of a typical location in a typical commercial or
hospital environment.
Note:
U
T
is the AC power voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic immunity - for equipment and systems that
are not LIFE-SUPPORTING
AccuScreen is intended for use in the electromagnetic environment specified below. The user of AccuScreen should ensure that it
is used in such an environment.
Immunity test
IEC 60601 test level
Complianc
e level
Electromagnetic environment - guidance
Radiated RF
IEC 61000-4-3
150 kHz to 80 MHz
outside ISM bands
a
3 V/m
80 MHz to 2.5 GHz
3 V/m
Portable and mobile RF communications equipment should be
used no closer to any part of AccuScreen, including cables, than
the recommended separation distance calculated from the equa-
tion applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2
d = 1.2
for 80 MHz to 800 MHz
d = 2.3
for 80 MHz to 2.5 GHz,
where
P
is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in metres (m).
b
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
c
should be less than the compliance
level in each frequency range.
d
Interference may occur in the vicinity of equipment
marked with this symbol.
Note 1:
At 80 MHz and 800 MHz the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Summary of Contents for AccuScreen ABR
Page 1: ...AccuScreen OAE ABR Screener User Manual Doc no 7 50 0920 US 02 Part no 7 50 09200 US 0459 ...
Page 8: ...8 Otometrics Table of Contents AccuScreen OAE ABR Screener User Manual ...
Page 66: ...66 Otometrics Printing Printouts from label printer ...
Page 82: ...82 Otometrics Handling Data in AccuScreen Deleting data ...
Page 88: ...88 Otometrics Maintenance and Cleaning Battery safety and maintenance ...
Page 98: ...98 Otometrics App 2 Status and Error Messages Usage and test related messages ...