Preparing for testing
Preparing the patient
4.4.2
Fitting the eartip on the probe
1. Select an ear tip that fits the patient's ear canal. You may have to try out a num-
ber of sizes in order to select the appropriate size.
2. Gently push the eartip onto the probe tip until it rests firmly against the base of
the probe. It is much easier to attach and remove the eartip if you turn it gently.
When you do so, make sure that you hold the probe by the probe body and
not
by the cable
.
Note ·
Accurate testing is only guaranteed if you use the eartips provided.
Note ·
The probe can be used for both ears. If you suspect infection in one ear, change
the eartip and clean the probe tip, or replace it with a spare, before you continue testing
on the other ear.
Warning ·
Using a probe with an unsuitably sized eartip or applying excessive force may
irritate the ear canal.
Note ·
In case of a probe error, make sure that the probe tip channels are clear (see
11.1 The probe tip and probe body
X
83
) and that the probe is connected.
See also
8.2 Probe test
X
67
.
4.4.3
Inserting the probe with eartip in the patient’s ear canal
1. When you have fitted an eartip on the probe, gently pull the pinna back and
slightly down and insert the probe in the ear canal, twisting the probe slightly as
you insert it.
The probe can be inserted with the probe cable pointing either upwards or
downwards, depending on which direction fits best.
Make sure that the probe fits well. Any leakage may increase the test duration
because of distortion and excessive noise.
Note ·
Make sure that the cable is not in contact with any vibrating surfaces during test-
ing.
Summary of Contents for AccuScreen ABR
Page 1: ...AccuScreen OAE ABR Screener User Manual Doc no 7 50 0920 US 02 Part no 7 50 09200 US 0459 ...
Page 8: ...8 Otometrics Table of Contents AccuScreen OAE ABR Screener User Manual ...
Page 66: ...66 Otometrics Printing Printouts from label printer ...
Page 82: ...82 Otometrics Handling Data in AccuScreen Deleting data ...
Page 88: ...88 Otometrics Maintenance and Cleaning Battery safety and maintenance ...
Page 98: ...98 Otometrics App 2 Status and Error Messages Usage and test related messages ...