Otometrics
51
Tests
The ABR test
5.5.2
Impedance testing and calibration
Testing both ears
• If you are testing with ear couplers, the
Bin-
aural
button is shown.
Press this button to test both ears.
Electrode impedance testing followed by calibra-
tion are performed before the actual test starts.
All three impedance and balance indicators
should prefereably be green (see
Impedance test
indicators
X
51
), but the test can be started even
if the indicators are yellow.
The
Start
button is greyed out until
Stimulus
and
Impedance
values are okay.
1. Press the
Start
button to start the test.
Impedance test indicators
Stimulus
The acoustical stimulus is being calibrated.
If the
Impedance
and
Balance
values are
OK
(green), press the
Start
button to start the test.
Red (nape)
The
Red
reading indicates the sum of the
Red
and the
Black
(ref-
erence) electrodes’ skin impedances.
White (high on the forehead)
The
White
reading indicates the sum of the
White
and the
Black
(reference) electrodes’ skin impedances.
Color coding
Summary of Contents for AccuScreen ABR
Page 1: ...AccuScreen OAE ABR Screener User Manual Doc no 7 50 0920 US 02 Part no 7 50 09200 US 0459 ...
Page 8: ...8 Otometrics Table of Contents AccuScreen OAE ABR Screener User Manual ...
Page 66: ...66 Otometrics Printing Printouts from label printer ...
Page 82: ...82 Otometrics Handling Data in AccuScreen Deleting data ...
Page 88: ...88 Otometrics Maintenance and Cleaning Battery safety and maintenance ...
Page 98: ...98 Otometrics App 2 Status and Error Messages Usage and test related messages ...