68
Otometrics
Quality Tests
Probe test
•
8.2.2 Performing the Probe Test
X
68
, simply by inserting the probe in the test
cavity of AccuScreen.
•
8.2.1 From the Quality Tests menu
X
68
8.2.1
From the Quality Tests menu
8.2.2
Performing the Probe Test
1. Press the
Probe
button on the
Quality Tests
screen.
2. Follow the instructions in
8.2.2 Performing
the Probe Test
X
68
.
1. Insert the probe tip without eartip in
the test cavity.
The test starts automatically.
Note ·
The Probe Test does not start auto-
matically e.g., when a patient is being cre-
ated.
If the probe functions correctly, the
message
Probe OK
appears.
If the probe does not function cor-
rectly, the message
Probe Test
failed!
appears.
You will also see a list of possible
errors.