Otometrics
85
Maintenance and Cleaning
Eartips
Calibration frequency
The probe should be calibrated annually by factory authorized service personnel.
11.2
Eartips
Eartips are disposable and should not be cleaned.
Note ·
Eartips should be disposed of according to local regulations.
11.3
Cleaning the ear coupler cable
Always mount the ear couplers on the ear coupler adapters before use to prevent the
ear coupler adapters from coming into contact with the patient. The ear couplers are
disposable items and should be used only on one patient.
If there is a risk of cross-infection, clean the adapters. Otherwise, clean the adapters
at the end of the day:
1. Disconnect the ear coupler cable from AccuScreen.
Caution ·
Never immerse the ear coupler cable in liquid.
2. Use a wet tissue with alcohol to clean the ear coupler cable and the ear coupler
adapter. Wipe only with disinfectants and make sure that liquid does not enter
the ear coupler adapter.
11.4
Cleaning AccuScreen
• Before cleaning, switch off AccuScreen and unplug it from any external power
source.
• Unplug the probe from AccuScreen.
Caution ·
Use only wiping disinfectant.
Caution ·
Always make sure that no moisture enters the probe, the sockets (data interface
connector and probe plugs) or the test cavity.
Caution ·
Never immerse AccuScreen into water or other cleaning solutions.
• Clean AccuScreen with a moist cloth - if required, use a little surface disinfec-
tant.
11.4.1
Cleaning and disinfecting the test cavity
If needed, use a wet tissue with alcohol to clean the entry surface of the test cavity.
Summary of Contents for AccuScreen ABR
Page 1: ...AccuScreen OAE ABR Screener User Manual Doc no 7 50 0920 US 02 Part no 7 50 09200 US 0459 ...
Page 8: ...8 Otometrics Table of Contents AccuScreen OAE ABR Screener User Manual ...
Page 66: ...66 Otometrics Printing Printouts from label printer ...
Page 82: ...82 Otometrics Handling Data in AccuScreen Deleting data ...
Page 88: ...88 Otometrics Maintenance and Cleaning Battery safety and maintenance ...
Page 98: ...98 Otometrics App 2 Status and Error Messages Usage and test related messages ...