12
6.1 Installing in 1" observation wells
Required accessories: Adapter ring for 1" observation wells.
Install the OTT Orpheus Mini as follows
Preliminary work: If not already done, install batteries
(see Chapter 5).
Open the observation well cover.
Determine the current depth with a contact gauge and make a note of it.
Slide O-ring from the communication unit onto the pressure probe cable.
Place adapter ring over the pressure probe cable and slide it until it comes
to rest against the communication unit.
Slide O-ring back onto the communication unit.
Feed the pressure probe into the observation well.
Slowly
and
carefully
lower the pressure probe with the pressure probe
cable!
Feed the communication unit into the observation well until the O-ring sits
on the observation well (see Figure 6).
Set operating parameters (see Chapter 7).
Close the observation well cover carefully.
Caution:
Do not damage the
infrared interface (maintain correct spacing)!
Fig. 6: Installing the OTT Orpheus Mini
in 1" observation well.
Adapter ring 1"
O-ring
Communication unit
Observation well
cover (example)
Infrared interface
1" observation well