19
– Setting up a communication link with the OTT IrDA-Link USB reading head:
Connect the OTT IrDA-Link USB reading head to a USB port on the PC. (USB
interface drivers must be installed – see separate installation instructions.)
Place the OTT IrDA-Link USB reading head on the OTT Orpheus Mini
infrared interface (see Fig. 12).
Select the communication path "IrDA – OTT IrDA-Link" in the OTT Orpheus
Mini operating program.
Notes
The integrated infrared interface of a PC (standard for many notebooks) can
only be used with the Windows 95 and Windows NT operating systems!
The OTT Orpheus Mini infrared interface has a radiation angle of approx. ±15 °.
Caution:
OTT Orpheus Mini connected to OTT ITC: Establishing a communi-
cation connection can take up to a minute! (This is always the case if the OTT
Orpheus Mini attempts to communicate with the OTT ITC (e.g. SMS message)
and the OTT ITC is open at the time.)
7.3 Adjusting the OTT Orpheus Mini operating parameters
How to set the OTT Orpheus Mini operating parameters
Click the "Connect" button (alternatively: menu "OTT Orpheus Mini", function
"Connect")
➝
the operating program reads the current operating parameters
of the OTT Orpheus Mini.
Adjusting operating parameters (see pages 21/22).
If necessary: "Set up cyclical data transfer using SMS messages in combination
with the intelligent top cap OTT ITC" or "Set up cyclical data transfer using
GPRS in combination with the intelligent top cap OTT ITC" (see page 20).
Fig. 12: Establishing a communication
link PC/OTT Orpheus Mini.