D.3
"Preset Single Register"
function (0x06)
1)
Do not write to this register
Register name
Register-
number Type
Value range
Fartory
setting
Access-
mode
Unit
䊳
RS-232 protocol
0011
unsigned word 1
➝
NMEA
R
[1]
䊳
RS-485 protocol
0012
unsigned word 1
➝
Modbus
2
➝
SDI-12
2
R
[1]
Register name
Register-
number Type
Value range
Factory
setting
Access
mode
Unit
䊳
Bus address
0000
unsigned word 1 … 255
1
W
–
䊳
Baud rate
0001
unsigned word 0
➝
9600
1
➝
38400
2
➝
57600
3
➝
115200
9600
W
[Bit/s]
䊳
–
1)
0002
䊳
Internal filter
type
0003
unsigned word 0
➝
IIR filter
1
➝
Floating mean
1
W
[1]
䊳
Filter length
0004
unsigned word 1, 16 … 256
1
W
[1]
䊳
Flow
direction setting
0005
unsigned word 0
➝
Both
1
➝
Towards sensor
2
➝
Away from sensor
0
W
[1]
䊳
Measuring-
sensitivity
0006
unsigned word 0 … 100
W
[1]
䊳
–
1)
0007
䊳
RS-232 protocol
0008
unsigned word 1
➝
NMEA
W
[1]
䊳
–
1)
0009
䊳
–
1)
000A
䊳
–
1)
000B
31