Troubleshooting
490E75=0_C
Ottobock | 129
Error message;
Error code
Content
Cause
Possible measure
L Ind Lamp
Short;
7206
Short circuit in electri-
cal circuit of left direc-
tion indicator
e.g. defective plug connection,
cable break
Defective direction indicator
signal
Check plug connections and
cable to direction indicator.
Replace lighting.
R Ind Lamp
Short;
720A
Short circuit in electri-
cal circuit of right direc-
tion indicator
L Ind Lamp
Failed;
7207
Error in electrical
circuit of left direction
indicator
e.g. defective plug connection,
cable break
Defective bulb
Check plug connections and
cable to direction indicator.
Replace lighting.
R Ind Lamp
Failed;
7208
Error in electrical
circuit of right direction
indicator
Over-current
Current in an actuator
circuit too great
e.g. defective plug connection,
cable break
Defective actuator or controller
Check plug connections and
cable to actuator. Verify that actu-
ating drive can move freely.
Replace actuator or controller.
Overtemp. (Acts) Temperature on actua-
tor circuits on controller
too high
e.g. defective plug connections
Defective actuator
Cool down phase
Check plug connections and
cable to actuator.
In case of recurrence: replace
actuator.
Overtemp.
(Lamps)
Temperature on the
light circuit on the cont-
roller too high
e.g. defective plug connections
Defective lamp
Cool down phase
Check plug connections and
cable to light.
In case of recurrence: replace
light.
Memory Error
Unspecified memory
error in control modu-
les
e.g. defective plug connections
Defective control module
Check and replace controller
with modules.
PM Memory
Error
Memory error in cont-
roller
e.g. defective plug connection,
cable break
Defective controller
Check all plug connections and
cables on controller.
Replace controller.
Bad Cable
Error on the communi-
cation cables between
control modules
e.g. defective plug connection,
cable break
Check all plug connections and
cables on the control modules
(incl. controller).
Replace cables.
Module Error
Error in control modu-
les
e.g. unspecified error in dis-
played control module
Displayed control module defec-
tive
Check all plug connections and
cables.
Replace controller.
System Error
System error
Non-attributable error in control
modules (primarily from third-
party providers)
Check all plug connections and
cables. Disconnect/test any mo-
dules from third-party providers
in succession if applicable.
Replace controller.
SID Detached
Special control discon-
nected
Special control (sip and puff
control) disconnected from LCD
monitor
Special control connection to
LCD monitor defective
Check all plug connections and
cables between special control
connection and sip and puff
control. Check programming
parameters.
Replace control panel.
Summary of Contents for 490E75 Series
Page 1: ...490E75 0_C Service manual...
Page 38: ...Service work 490E75 0_C 38 Ottobock 57 58 59 60 61 62 63...
Page 41: ...Service work 490E75 0_C Ottobock 41 70 71 72 73 74 tt 75 76...
Page 55: ...Service work 490E75 0_C Ottobock 55 103 104 105 106 107 108 109 110...
Page 108: ...Service work 490E75 0_C 108 Ottobock 227 228 229 t 230 231 232...
Page 112: ...Service work 490E75 0_C 112 Ottobock 237 238 239 240 241...
Page 143: ...R Net wheelchair control 490E75 0_C Ottobock 143 Seat module pin assignment 287...
Page 161: ...490E75 0_C Ottobock 161...
Page 162: ...490E75 0_C 162 Ottobock...