R-Net wheelchair control
490E75=0_C
154 | Ottobock
Parameter
Setting range Description
Standby Right
Drive, seat,
Bluetooth,
external de-
vices, mode 5,
mode 6, IR
mode, pro-
gramming
Used to set which mode is switched to if the joystick is moved to the right
when the system is in standby mode. Please note that Mode select in standby
must be set to Yes in order for this function to be active.
Remote Selec-
tion
Yes, no
Used to set whether profiles or modes can be selected and input using joy-
stick movements.
Setting the parameter to Yes makes it possible to switch from a profile into
the remote selection mode. The remote selection mode can be reached via
standby mode, the profile button or an external input device for the profile.
The profile can also be selected from the remote selection mode.
If the parameter is set to No, it is not possible to access the remote selection
mode from the profile and the profile cannot be selected from the remote
selection mode either.
Background
Blue, white
Background colour of the LCD display for the individual profiles. Standard
setting is blue.
7.4.5 Latched drive parameters
The following parameters can be set under Latched:
Parameter
Setting range Description
Latched Drive
Off, Step, Step
reverse, Cruise,
Cruise reverse
Type of latched drive mode:
• Off: Latched drive mode is switched off.
• Step: Step mode in the forward direction only. Driving in step mode me-
ans that the latched speed of the wheelchair is increased or decreased
using short joystick movements.
• Step reverse: Step mode in both the forward and reverse direction
• Cruise: Cruise mode in the forward direction only. Joystick movement
accelerates the wheelchair and driving continues at the speed reached
when the joystick is released.
• Cruise reverse: Cruise mode in both the forward and reverse direction.
If Latched drive is set in a profile, an emergency stop switch must be con-
nected to the socket for the external input device for the profile on the joystick
module.
Latched Actua-
tors
Yes, no
Operation of seat functions in latched mode, i.e. the seat function is trigge-
red with a short joystick movement. If Latched actuators is set in a profile, an
emergency stop switch must be connected to the socket for the external input
device for the profile on the joystick module.
Latched Timeout 0 to 250 s in
single steps
Timeout for driving and operation of the variable motors in latched mode.
The timeout period is the maximum period of time for which the joystick can
remain in the neutral position before movement in latched mode is ended. If
the period of time is exceeded, movement in latched mode stops. Timeout
is therefore an important safety function with which driving or movement of
a variable motor is stopped when the user is unable to stop the functions as
usual. It is important that the correct duration is set for the timeout.
When a value of 2 or lower is set, the timeout function is disabled. This
should only take place in exceptional cases and following a comprehensive
risk analysis.
Latched Timeout
Beep
Yes, no
A beep is emitted shortly before expiration of the timeout period.
Summary of Contents for 490E75 Series
Page 1: ...490E75 0_C Service manual...
Page 38: ...Service work 490E75 0_C 38 Ottobock 57 58 59 60 61 62 63...
Page 41: ...Service work 490E75 0_C Ottobock 41 70 71 72 73 74 tt 75 76...
Page 55: ...Service work 490E75 0_C Ottobock 55 103 104 105 106 107 108 109 110...
Page 108: ...Service work 490E75 0_C 108 Ottobock 227 228 229 t 230 231 232...
Page 112: ...Service work 490E75 0_C 112 Ottobock 237 238 239 240 241...
Page 143: ...R Net wheelchair control 490E75 0_C Ottobock 143 Seat module pin assignment 287...
Page 161: ...490E75 0_C Ottobock 161...
Page 162: ...490E75 0_C 162 Ottobock...