03/2010
B500 | B500-S
Page 123
Operation
ATTEnTIOn !
Unauthorized battery replacement!
The battery may only be replaced by a spe-
cialist dealer. The characteristic curve of
the battery charger set at the factory corre-
sponds to the battery provided and must not
be changed. Setting the characteristic curve
incorrectly can result in permanent damage to
the battery.
The battery type can be set by means of a switch located
at the bottom of the battery charger (covered by a green
adhesive label). Once the battery type has been set, the
green LED flashes at the beginning of the charging proc-
ess.
When handling the battery charger, the following safety
instructions must be observed:
r
Always place the rubber feet of the battery charger
on a level surface.
r
Protect the battery charger from direct sunlight to
prevent the charger from getting any warmer than it
already is.
r
The place where the battery charger will be used must
be dry and well ventilated. Avoid letting dust and dirt
enter and affect the function of the battery charger.
r
To clean the charger, use a dry piece of cloth.
DAnGEr !
Explosion hazard as a result of sparks!
Always switch off the battery charger and re-
move the mains plug before disconnecting the
battery.
To charge the batteries, proceed as follows:
1. Turn off the controls on the power wheelchair.
2. Connect the plug of the battery charger to the charg-
ing receptacle on the control panel of the power
wheelchair (see section 7.3.1, fig. 13, Pos.10).
3. Connect the battery charger to a wall socket and turn
it on. The batteries will start charging automatically,
and the current charge status is indicated on the LEDs
of the battery charger (see table 6).
4. When the charging process is complete, turn off the
battery charger: Disconnect the battery charger from
the mains supply, then from the control panel.
5. Turn on the controls, and the power wheelchair is
ready for driving.