22
Bluetooth ID number
13.4 Directives and manufacturer’s declaration
13.4.1 Electromagnetic environment
This product is designed for operation in the following electromagnetic environments:
•
Operation in a professional healthcare facility (e.g. hospital, etc.)
•
Operation in areas of home healthcare (e.g. use at home, use outdoors)
Observe the safety notices in the section "Information on proximity to certain areas" (Information
on Proximity to Certain Areas).
Electromagnetic emissions
Interference meas
urements
Compliance
Electromagnetic environment directive
HF emissions accord
ing to CISPR 11
Group 1/class B
The product uses HF energy exclusively for its
internal functioning. Its HF emissions are
therefore very low, and interference with neigh
bouring electronic devices is unlikely.
Harmonics according
to IEC 61000-3-2
Not applicable –
power below 75 W
–
Voltage fluctu
ations/flicker accord
ing to IEC 61000-3-3
Product meets the
requirements of the
standard.
–
Electromagnetic interference immunity
Phenomenon
EMC basic standard
or
test procedure
Interference immunity test level
Electrostatic dis
charge
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
High-frequency elec
tromagnetic fields
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
Magnetic fields with
rated power frequen
cies
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
Electrical fast transi
ents/bursts
IEC 61000-4-4
± 2 kV
100 kHz repetition rate
Surges
Line against line
IEC 61000-4-5
± 0.5 kV, ± 1 kV
Conducted interfer
ence induced by high-
frequency fields
IEC 61000-4-6
3 V
0.15 MHz to 80 MHz
6 V in ISM and amateur frequency bands
between 0.15 MHz and 80 MHz
80% AM at 1 kHz
Voltage drops
IEC 61000-4-11
0% U
T
; 1/2 period
At 0, 45, 90, 135, 180, 225, 270 and 315
degrees