B-526-10N-E
40
Parameter
Code
Default Setting
Description
Remarks
Pulse width
Pon
1 or 50
"An item shown on
parameter label"
Factored pulse output "ON"
width
(Unit: [msec])
Ex.: To change pulse width from 1 ms to 50ms
→
Set to "Pon 50" (msec).
(Note 4)
Pulse dummy
output 1
Pd1
Not a parameter to
be set up
Furnishes a 1 Hz simulated
factored pulse output
irrespective of flowmetering.
・
This function is useful in loop check, etc.
・
For operation in practice, see "About Dummy
Output Functions (special features)."
Pulse dummy
output 2
Pd2
Not a parameter to
be set up
Furnishes a 10Hz simulated
factored pulse output
irrespective of flowmetering.
NOTES 1. When conversion factor (H) has been set, change pulse weight (Po), indicated unit (Un), etc. must be changed relative to the
units after conversion.
2. Be sure to set up a value such that
F
×
H
2
≦
Pu
≦
F
×
H
×
10000
3. Not shown on the LCD if the meter operates on the battery alone.
4. Be sure to set a value such that factored pulse "OFF" width > 1 ms.
※
To see the factory-default setting parameters upon delivery, please refer to the "Parameter Settings Table" that is shipped
with the product.