13. SENSOR REPLACEMENT PROCEDURE
①
Take off four hex socket head screws on the side
of the register.
SENSOR UNIT
REGISTER
ASSEMBLY
SCREWS
②
Carefully draw the register assembly out. Exercising
care not to bump the sensor against adjacent
components, draw it out in the horizontal direction.
REGISTER ASSEMBLY
③
Take off four hex socket head bolts and separate
the cover.
ELECTORONICS UNIT
④
With screwdriver, take off two fitting screws
holding the electronics unit.
ELECTRONICS UNIT
⑤
Holding the electronics unit in both hands,
carefully draw it out
CONNECTOR
⑥
Uncouple the connector from the sensor unit at
back of the electronics unit.
C-SHAPED
STOP RING
CONNECTOR
⑦
Using C-shaped stop ring pliers, remove the
C-shaped stop ring for the shaft. The sensor unit
is now separable from the register housing.
SENSOR RETAINER
FITTING SCREW
SENSOR BRACKET
⑧
Loosen the sensor retainer fitting screws (M4)
with screwdriver, remove the sensor retainer and
draw out the sensor unit.
BLIND
PLUG
SENSOR
HOLE
SENSOR BRACKET
Location of Hole to Install the Sensor
SENSOR BRACKET
SENSOR
153
152
158
⑨
Install a new sensor unit through the opening
through which the old sensor unit was removed
and assemble in the reverse order of disassembly.
Summary of Contents for ULTRA UF-II
Page 67: ...B 526 10N E 67...