B-526-10N-E
59
17. GENERAL SPECIFICATIONS
17.1 Standard ULTRA register Specification
4. Details of the explosionproof construction comply with 94/9/EC Directive.
Type
Exd II BT4
Equipment Group
II
Equipment Category
2
Type of Explosive Atmosphere
G
Applicable Hazardous Area
Zone 1 and Zone 2
●
Indicated Output Units
Meter
Size
Total Counter Units (pulse unit)
L
Instantaneous Flowrate Units
bi: Hourly
b2: Per minute
82
1 (standard), 10,100
b1 =
×
100 L/h
b2 =
×
1L/min
83
1 (standard), 10,100
b1 =
×
100 L/h
b2 =
×
1L/min
84
1 (standard), 10,100
b1 =
×
100 L/h
b2 =
×
1L/min
85
10 (standard), 100,1000
b1 =
×
1kL/h
b2 =
×
10L/min
86
10 (standard), 100,1000
b1 =
×
1kL/h
b2 =
×
10L/min
87
10 (standard), 100,1000
b1 =
×
1kL/h
b2 =
×
10L/min
88
10 (standard), 100,1000
b1 =
×
1kL/h
b2 =
×
10L/min
NOTE: Total flow units remain the same as pulse units.
Item
Description
Meter Size
82, 83, 84
※
85, 86, 87, 88
※
Local
display
(LCD)
(Note 1)
Grand total (8-digit) units
0.001m
3
(standard), 0.01 m
3
, 0.1m
3
0.01m
3
(standard), 0.1 m
3
, 1m
3
Resettable counter
(7-digit)
C mode
Instantaneous
flowrate (4-digit)
b1 mode
0.1m
3
/h
1m
3
/h
b2 mode
0.001m
3
/min
0.01m
3
/min
O
u
tp
u
t
-
None
Local display only
C
u
rr
e
n
t
Analog
4 to 20mA DC See Acceptable load resistance range on page 19.
Pulse
(Note 2 and 3)
Category
Factored or unfactored: 0/1 = 4/20mADC
Pulse width
Factored: 1 ms (standard), 50 ms Unfactored: 2 ms
Voltage pulse unit
Same as the indicated unit
O
pe
n
Pulse
(Note 2 and 3)
Category
Factored or unfactored: Max. voltage impression 30VDC Allowable current 50mA
Pulse width
Factored: 1 ms (standard), 50 ms Unfactored: 2 ms
Voltage pulse unit
Same as the indicated unit
Vo
lta
g
e
Pulse
(Note 2 and 3)
Category
Factored or unfactored: 0/1 = 1 VDC max. / 7V DC min.
Pulse width
Factored: 1 ms (standard), 50 ms Unfactored: 2 ms
Voltage pulse unit
Same as the indicated unit
Power
Supply
Output not provided
Built-in dedicated lithium battery pack good for 8 years approx.
(Depends on operating conditions. 2 years with explosionproof constraction
④
and
⑤
.)
Output provided
External power source 12 to 45V DC (analog, current pulse)
12 to 24V DC (open collector pulse, voltage pulse)
12 to 45V DC (analog, current pulse combined)
Current drain: 30mADC max. See the acceptable load resistance range on page 19. (Note 1)
Signal Transmission Cable
Externally shielded cabtyre cable (VCTF 1.25mm
2
) Finished O.D. 8.5 to 12 mm (Note 5)
Transmission Length
One kilometer max.
Transmission Lines
2-wire system
Analog or current pulse
3-wire system
Open collector pulse or voltage pulse
4-wire system
Analog P current pulse
Ambient Temperature
-10 to + 60
°
C
Explosionproof Construction
Select either one of the following:
①
Non-explosionproof type
②
TIIS
:
Exd
Ⅱ
BT4/Exia
Ⅱ
BT4
③
NEPSI
:
Exd
Ⅱ
BT4
④
ATEX
:Ⅱ
2GExd
Ⅱ
BT4
⑤
FM
:
Class l.Division l/Group C,D T4 AEx/Exd
Ⅱ
BT4
⑥
KOSHA
:
Exd
Ⅱ
BT4
Applicable EU Directives
RoHS
:
2011/65/EU
EMC
:
2014/30/EU
ATEX
:
94/9/EC (Note 4)
Applicable EN standard
RoHS
:
EN50581
:
2012
EMC
:
EN61326-1
:
2013 Class A
ATEX
:
EN60079-0
:
2006, EN60079-1
:
2007
Enclosure Protection Rating
1P66 (dust tight /watertight) IEC / EN60529, JIS C 0920
Enclosure Material
Aluminum die cast
Finish
Munsell2.5PB 5/8 (baked melamine resin coating
NOTES: 1. When powered by the battery pack, the register provides local readout only with no remote output.
2. If pulse width in excess of 1ms is desired with minimum factored pulse unit, the max. flowrate
can possibly be restricted; consult the factory.
3. With meter size marked
※
, if minimum factored pulse unit is selected, pulse width other than 1
ms is unacceptable.
5. For wiring of explosionproof type (output available models), do not fail to use the ancillary pressure-
resistant packing.
Also, in case of TIIS explosionproof type used under the ambient temperature of 45
℃
or higher,
use a cable resistant to the temperature of 75
℃
or higher.