B-526-10N-E
60
●
Full Scale Frequencies of Unfactored Pulses
Meter Size
Min. Flowrate, kL/h/Frequency, Hz Max. Flowrate, kL/h/Frequency, Hz
82
8.5/4.9
300/171.2
83
12/4.1
410/141.7
84
25/5.0
590/117.9
85
45/4.4
950/192.9
86
65/3.5
1400/
1
74.9
87
110/3.5
2000/
1
63.3
88
170/3.1
2800/
1
50.4
17.2 Smart Type ULTRA register Specification
Item
Description
Meter Size
82, 83, 84
※
85, 86, 87, 88
※
Local
display
(LCD)
Mode
select
Grand total
(8-digit)
-
×
1L (std.),
×
10L,
×
100L
×
10L (std.),
×
IDOL,
×
1000L
Instantaneous
Flowrate
(4-digit)
b mode
×
100 L/h
×
1 m
3
/h
p mode
0 to 100 %
Bar graph
8-segment bar graph
O
ut
p
ut
C
ur
re
nt
Analog
4 to 20mADC Damping: 0 to 100 sec. See Acceptable load resistance
range on page 19.
Pulse
(Note 4 and 5)
Category
Factored or unfactored: 0/1 = 4/20mADC (Note 1)
Pulse width
Factored: 1 to 1000 ms (default 1 ms), Unfactored: 2 ms
Factored pulse unit
Same as the indicated unit
O
p
en
co
lle
ct
o
r
Pulse
(Note 4 and 5)
Category
Factored or unfactored: Max. voltage impression 30VDC Allowable current 50mA (Note 1)
Pulse width
Factored: 1 to 1000 ms (default 1 ms), Unfactored: 2 ms
Factored pulse unit
Same as the indicated unit
Alarm output status
Max. allowable voltage impression: 30VDC Allowable current: 50mA
Communication
Current output 4 to 20mA Instant flowrate signal superimposed on BELL 202 FSK (freq. modulated)
Power Source
12 to 45VDC Current drain 30mA max. See Acceptable load resistance range on page 19
Ambient Temperature
-
10 to
+
60
℃
Functions
3 a correction
Corrects for volumetric expansion coefficient of flowmeter's measuring chamber.
Meter error correction
A 4-point broken line correction. Meter error at four flowrate points is
entered in % meter error.
Instant flowrate span setting
Any point more than 3 times the minimum flowrate and less than twice the
maximum flowrate within the accuracy guaranteed range.
4 - 20mADC trim
Communications
Respective parameters setting and reading with EL2310. Communicatin under HART protocol.
Alarm
Normally transistor OFF; in error, transistor ON (Tells the operator sensor
error, full scale exceeded, or preset value exceeded.)
(Note 3)
Self test loop check
Accuracy
Total flow
±
0.01%
±
1 count
Instantaneous flowrate
±
0.3% of F.S.
Signal Transmission Cable
Externally shielded cabtyre cable (VCTF 1.25mm2) Finished O.D. 8.5 to 12mm (Note 2)
Transmission Length
One kilometer max.
Transmission Lines
2-wire system
Analog or current pulse
3-wiresystem
Analog or current pulse + open collector
Construction
Select one of the following enclosures:
①
Non-explosionproof type
②
Explosionproof type: TIIS Exd
Ⅱ
BT4/Exia
Ⅱ
BT4
③
Explosionproof type: KOSHA Exd
Ⅱ
BT4
Finish
Munsell 2.5PB 5/8
NOTES: 1. Current pulse and open collector pulse type are of the identical setting.
A combination factored and unfactored output is not available.
2. For wiring of explosionproof type, do not fail to use the ancillary pressure-resistant packing.
Also, in case of TIIS explosionproof type used under the ambient temperature of 50
℃
or higher,
use a cable resistant to the temperature of 70
℃
or higher.
3. In the event of an alarmed condition, the display shows an error message and flow information
alternately.
4. If pulse width in excess of 1ms is desired with minimum factored pulse unit, the max. flowrate
can possibly be restricted; consult the factory.
5. With meter size marked
※
, if minimum factored pulse unit is selected, pulse width other than
1ms is unacceptable.