Thermo Scientific

Section 1

General Information

The EC-I05 Power Supply is intended for use with any
electrophoretic device designed to operate below 250 Volts and 500
milliAmps DC. 

Two sets of outputs operate in parallel to provide between 20 and 250
volts, up to 500 milliamps, and up to 75 watts in the constant voltage
mode. 

A single LED display provides output voltage or current readings. 

Overcurrent protection is provided by audible and visual alarms and a
single 1 amp slo blow fuse on the main circuit. 

Intended Use and

Specifications

Summary of Contents for EC-105

Page 1: ...Electrophoresis Power Supply EC 105 Operation and Maintenance Manual 7218471 Rev 0 Visit us online to register your warranty www thermoscientific com warranty ...

Page 2: ...Thermo Scientific MANUAL NUMBER 7218471 0 5 2 12 Transfer to Marietta was 157 3000 00 E ccs REV ECR ECN DATE DESCRIPTION By Preface EC 105 i ...

Page 3: ... and maintenance must be performed by qualified service personnel s Material in this manual is for information purposes only The contents and the product it describes are subject to change without notice Thermo Fisher Scientific makes no representations or warranties with respect to this manual In no event shall Thermo be held liable for any damages direct or incidental arising out of or related t...

Page 4: ...ment that has been put on the market after 13 August 2005 This product is required to comply with the European Union s Waste Electrical Electronic Equipment WEEE Directive 2002 96 EC It is marked with the WEEE symbol Thermo Fisher Scientific has contracted with one or more recycling disposal companies in each EU Member State European Country and this product should be disposed of or recycled throu...

Page 5: ...l your needs for spare or replacement parts or provide you with on site service We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products Whatever Thermo Scientific products you need or use we will be happy to discuss your applications If you are experiencing technical problems working together we will help you locate the problem and chances are correct ...

Page 6: ...entific Table of Contents General Information 1 1 Intended Use and Specifications 1 2 Set Up 2 1 Operation 3 1 Precautions and Service 4 1 Cleaning 4 1 Technical Specifications 4 2 Section 1 Section 2 Section 3 Section 4 ...

Page 7: ......

Page 8: ...ermo Scientific Section 1 General Information Figure 1 1 EC 105 Components 1 Main Power Switch 2 Check Fuse LED 3 Voltage Selec Control 4 Low High Range Switch 5 Overcurrent LED 6 Volts Milliamp Display 1 2 3 4 5 6 ...

Page 9: ... 500 milliAmps DC Two sets of outputs operate in parallel to provide between 20 and 250 volts up to 500 milliamps and up to 75 watts in the constant voltage mode A single LED display provides output voltage or current readings Overcurrent protection is provided by audible and visual alarms and a single 1 amp slo blow fuse on the main circuit Intended Use and Specifications ...

Page 10: ...ltage is within the stated range The current rating is within the capacity of your supply outlet Warning This equipment must be earthed s The wires in the electrical lead are colored in accordance with the following code Green Earth Ground White Neutral Common Black Live Phase Connect the wires to a non reversible plug as follows Green wire to the terminal marked E Earth G Ground or colored Green ...

Page 11: ...wer supply in proximity to the electrophoretic cell s with which it is to be connected Be sure to place the power supply in a safe dry location with all the controls accessible Connect to Cell and Mains Connect power supply to electrophoretic cell then to the AC receptacle ...

Page 12: ... incremental adjustment The Lo Hi range switch is set to the Hi position whenever operating voltages above 140 volts are required Note The overcurrent alarm will sound whenever maximum output current 500mA is exceeded Should this occur reduce the operating voltage until the overcurrent alarm goes off The volts milliamps display switch will display either volts to nearest volt or milliamps to the n...

Page 13: ......

Page 14: ...s for the unit to temperature equilibrate This power supply is equipped with a slo blow 1 amp fuse in the mains circuit Should this fuse open the check fuse LED on the front panel will illuminate Caution Use only 1 amp 3 4 amp for 240VAC mains slo blow fuses in this equipment s Should this product become excessively warm smoke or display erratic information during operation discontinue use immedia...

Page 15: ...40 150 160 170 180 190 200 210 220 230 240 250 Maximum Voltage 250 Volts Maximum Current 300 milliAmps 150 250 Volts 500 milliAmps below 150 Volts Maximum Power 75 Watts Number of Outputs 2 Fuse Replacement 1 Amp slo blow Temperature Range 0 C 40 C non condensing atmosphere Dimensions 8 W x 6 D x 4 5 H 20 3 cm x 13 3 cm x 11 4 cm Weight net 5 lbs 2 3 kg Weight shipping 7 lbs 3 2 kg Technical Speci...

Page 16: ... the original warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo Fisher Scientific postage paid and replacement parts are shipped FOB desti nation THIS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER WARRANTIES WHETHER WRITTEN ORAL OR IMPLIED NO WARRANTIES OF MERC...

Page 17: ... excluded from this warranty Replacement or repair of components parts or equipment under this warranty shall not extend the warranty to either the equipment or to the component part beyond the original warranty period The Technical Services Department must give prior approval for return of any components or equipment At Thermo s option all non conforming parts must be returned to Thermo postage p...

Page 18: ...Thermo Fisher Scientific 401 Millcreek Road Marietta Ohio 45750 United States www thermofisher com ...

Reviews: