5.Front Panel Operation
33
please note that this signal will not turn Level Low until the end of the last period (If
the Waveform has a relative starting phase, it may be not zero intersections).
To Set the DC Output
(1)
Press
Utility
and choose
Output Setup
, press
F3
to select DC.
(2)
Press
F3
to switch between On/Off.
(3)
If On is selected, you can set the DC voltage. Turn the
knob
to change the value; or
press the number keys to input the desired value, press
F1
or
F2
to select
the unit.
Current Parameter
DC Voltage
Figure 5-21: The DC Setting Interface
To Set the I/O
To Set the Baud Rate
Press
Utility
and choose
I/O Setup
, press
F1
to choose BaudRate, press
F1
to
switch the RS232 baud rate.
Make sure that the Baud Rate matches that of the computer. The optional rates are:
115.2K, 56K, 38.4K, 9.6K and 4.8K. The default setting is 115.2K.
To Set the LAN
Press
Utility
and choose
I/O Setup
, press
F2
to choose LAN, enter the submenu.
Press
F1
to set IP Addr, the range of each byte is 1 - 255.
Press
F2
to set Port, the range is 1 - 4000.
Press
F3
to set GateWay, the range of each byte is 1 - 255.
Setting method
: Turn the
knob
to change the value of cursor position in the Parameter
bar. Press
/
direction key to move the cursor.
After setting, wait a few seconds (saving the parameters requires some time), and restart
the waveform generator so that the new setting can be applied.
About the application of this setting, please refer to "