4.Quick Start
(3)
Set the output voltage to the maximum rating.
(4)
Turn on the channel output. The
ON/OFF
and
Curr/CC
key is lighted, which
indicates the channel you opened is in Constant Current output mode.
(5)
In each working mode, set some different current values; check if the actual current
value displayed is close to the set current value, and to check if the actual voltage
value displayed is nearly zero.
(6)
Check that if the output current can be adjusted from zero to the maximum rating.
(7)
Turn off the channel output and remove the short wire from the output terminals.
4.5
Working Mode
ODP Series are designed with four working modes:
Independent
,
Parallel
,
Series
and
Plus-minus
. Press the
Mode
key to switch between the four working modes. For the
instructions of the user interfaces in the four working mode, please see "
User Interface
"
on P7.
4.5.1
Icons and Ratings
The status icons and voltage/current ratings of the four mode is listed below.
Independent
Parallel
Series
Plus-minus
Status Icons
Voltage rating 0
~
30V 0
~
30V 0
~
60V 0
~±
30V
ODP3032
Current ratings
0
~
3A 0.1
~
6A 0
~
3A 0
~
3A
ODP3052
Current ratings
0
~
5A 0.1
~
10A 0
~
5A 0
~
5A
The 5V output terminals always output fixed 5V during the instrument is powered on, the
max output current is 3A (ODP3032) or 5A (ODP3052). Users could choose the proper
mode and output terminals as required.
4.5.2
Connections
Take ODP3032 for instance:
Independent
13