background image

28

reset

the

short-circuit

protection

parameters.

Over-Load P

The

software

detects an overload
error

Eliminate overload error or adjust overload
protection parameters.

Over Temperature

The

software

detects an over-
Temperature error

Check whether the power supply air duct
is blocked.

Error Resume

Automatic

error

recovery is enabled,
recoverable

errors

are detected, and
recovery

attempts

fail for 10 times

After confirming the cause of the error and
troubleshooting, restart the machine. The
error alarm can be cleared by pressing the
"EXIT" key on the main UI.

key is locked

Warning

Key locked

Press the “LOCK” key to unlock it.

Return to HOME

Operation method in
the home

Return to the main UI and operate.

Close Output

Method of operation
in

closed

output

state

Operation after closing output.

RemoteCntr:Comms

Operate

keys

in

remote mode

Press “L/R” to switch back to local control.

RemoteCntr:Analog
Switching Prior

Cannot start output
during

priority

switching

Open output later.

Please Later

The priority cannot
be switched again
during

priority

switching

Wait 1 second and switch the priority
again.

Step Mode Is En

Cannot enable other
mode in step mode

Operation after Turn off Step mode.

Chg Mode Is En

Cannot enable other
mode

in

charge

mode

Turn off charging mode before operation.

Func Mode Is En

Cannot Enable other
mode In function
generator mode

Turn off function generator before
operation.

Exit Setting

Illegal operation

Operation after exiting the Settings.

Invalid Operate

The save and call
functions

are

unavailable in the
current UI

Perform operations on the correct UI.

Value Exceeds

The

input

value

exceeds the legal

Input valid value.

Value Too Small

Summary of Contents for OWP H Series

Page 1: ...OWP_H Series User manual For product support visit www owon com hk download ...

Page 2: ...that in the materials published originally The information in this manual was correct at the time of printing However LILLIPUT will continue to improve products and reserves the rights to change specification at any time without notice is the registered trademark of the LILLIPUT Company Fujian LILLIPUT Optoelectronics Technology Co Ltd No 19 Heming Road Lantian Industrial Zone Zhangzhou 363005 P R...

Page 3: ... copy of the customers proof of purchase is also required This warranty shall not apply to any defect failure or damage caused by improper use or improper or inadequate maintenance and care We shall not be obligated to furnish service under this warranty a to repair damage resulting from attempts by personnel other than our company representatives to install repair or service the product b to repa...

Page 4: ...tion Generator Mode 10 3 2 Information 12 3 2 1 Error Log 12 3 2 2 Operating Log 12 3 2 3 Event Log 12 3 3 System Setting 13 3 3 1 UI Setting 13 3 4 User Setting 14 3 4 1 Communication Setting 14 3 4 2 Function Setting 14 3 4 3 Protect Setting 18 3 4 4 Password Setting 20 3 4 5 Reset 20 4 Technical Specifications 21 4 1 Product parameters 21 4 2 Product selection 22 5 Appendix 23 5 1 Accessory 23 ...

Page 5: ... 10 Table 8 Rectangular pulse trapezoidal wave generator 11 Table 9 Line generator 11 Table 10 Information 12 Table 11 Error log 12 Table 12 Operating log 12 Table 13 Event log 12 Table 14 System setting 13 Table 15 UI setting 13 Table 16 Communication setting 14 Table 17 Function setting 14 Table 18 Output timing setting 15 Table 19 Parallel connection 16 Table 20 Interface Setting 16 Table 21 Dr...

Page 6: ...tching analog1 3 Normally closed contact Output of the dry contact 3 Matching analog1 4 NC 4 Matching analog1 5 Input of the dry contact 5 Input of the dry contact 6 6 RS485 1 485 A CAN 2 CAN L 2 485 B 7 CAN H 3 9 NC 1 3 6 8 NC Table 1 Defines of interface Digital IO interface PIN 1 to 3 is a dry contact output interface with complementary functions of normally open and normally close PIN2 is the ...

Page 7: ... as the positive and negative of analog input 1 the positive and negative of analog input 2 the positive and negative of analog output 1 the positive and negative of analog output 2 If you need analog function please inform us of the specific requirements in advance 1 2 Voltage compensation Figure 3 Schematic diagram of voltage compensation wiring To use the remote voltage compensation function us...

Page 8: ...in series parallel each device will display master slave machine number and CAN data receiving and receiving status of the machine gray Note 1 display elements of the status area can be hidden When an application mode is enabled the status of the application mode will be displayed and when the dry contact is used the corresponding status icon will be displayed 2 Output mode include common mode and...

Page 9: ...time output information Waveform Home displays output real time information in an intuitive waveform manner Note 1 The Home is the only interface for setting voltage current and power reference values 2 Press ENTER to set the sampling rate of the waveform displayed on the waveform Home Whether the voltage current and power waveform are displayed can be controlled by pressing VOLT CURR or POWER key...

Page 10: ...EXIT Returns the previous level or exit setting 3 Press 3 times Power set SAVE Save current settings Clockwise rotation Increase value RECALL Recall the saved settings Up shift L R local remote control mode Anti Clock rotation Reduce value CONF Function Menu Down shift Table 2 Key description The operation area includes setting area function area digital area and knob See Appendix 1 for key detail...

Page 11: ...ed data and press ENTER key to confirm the callback data 2 On the application mode setting UI press RECALL key to bring up the relevant recall UI press or key to select the pre called data and press ENTER key to confirm the callback data Local Remote mode switching Press L R to switch the local remote mode temporarily in the Home auxiliary Home for temporary test the mode is not saved Buzzer contr...

Page 12: ...7 3 Menu Figure 9 Menu ...

Page 13: ...g curve Function generator mode This mode generates a variety of regular waveforms such as sine triangular zigzagged rectangular pulse trapezoidal and Line segment and a combination of these waveforms and superpositions the modified waveforms onto the DC Voltage or current output The preset function will provide all necessary parameters for the user such as baseline cycle number waveform amplitude...

Page 14: ...only Figure 10 Step Mode 3 1 2 Charge Mode Charging Mode Disable Num Of Steps 3 Steps Step Delay XXXXX mS Dry Contact Control Disable Num Of Cycles Infinite Step1 Volt Reference XX XXX V Curr Reference XX XXX A Judgment Volt XX XXX V Judgment Curr XX XXX A Charging Curr XXXXX S Charging Time XXXXX S Table 5 Charge Mode Charging mode parameters Charging mode Whether to enable charging mode Step num...

Page 15: ... dry contact action When the discharge period ends close the dry contact and go to the next step 2 Charging time is the time from voltage and current judgment logic of each step to discharge time float charging time 3 Discharge time is the time from the completion of each charge time to the next step during which the output will be closed Generally used for external discharge of energy storage med...

Page 16: ...XXX S Interval Time XXXXX S Table 8 Rectangular pulse trapezoidal wave generator Superimpose a rectangular pulsed or trapezoidal wave on a DC reference Wave Line Wave Ref CV Periods Infinite Parameter Of Wave Amplitude XX Segments Interval Time XXXXX S Parameter Of Segments Table 9 Line generator Output line wave by multiple line segments ...

Page 17: ...lt date and fault time 3 2 2 Operating Log 2 Records The Current Page 0 1 Norm 10kwh Time 20y 12m 12d 12h 1 CvSt 300wh Time 20y 12m12d 10h Table 12 Operating log The device automatically records each running information and you can query on the operating log page Each operating information contains running mode information output energy and scrolling date time and time information in turn 3 2 3 Ev...

Page 18: ...2 Theme Bar 3 Front Regular 4 Status Icon Show 5 Data Time Show 6 Auxiliary Home Show 7 Dim Screen Time 600 S Table 15 UI setting UI setting parameters Effect Settings Background color theme and font options When the theme is set as the progress bar the voltage current and power display bar will be displayed in proportion to the actual value and rated value Element display State icon buzzer state ...

Page 19: ...us address address range 1 247 default address is 1 3 4 2 Function Setting 1 Startup Mode Local 2 Rise Time Of Volt XXXXX mS 3 Fall Time Of Volt XXXXX mS 4 Rise Time Of Curr XXXXX mS 5 Fall Time Of Curr XXXXX mS 6 Auto Reco Fault 30 S Close 7 Auto Output Gold 30 S Close 8 Timing Output 9 Parallel Series Connection 11 Interface Setting Table 17 Function setting Function setting parameters Startup m...

Page 20: ...ttempt to recover 3 Auto output hold function is generally used in unattended scenarios When the power grid is off the device will remember the output state before the power grid is off and actively control the output after the program sets the time after the power grid recovers Timing Output 1 Time Reference Clock The First Group Disable Power On Time 08 h 00 m 00 s Power Off Time 10 h 00 m 00 s ...

Page 21: ...unit and the motion logic can be set to local or remote communication logic to control Association logic When control mode is set to Local the dry contact can be associated controlled by Fault Power on off Condition or time logic Signal delay Dry contact action delay time Note After the dry contact output function is enabled the corresponding status of the dry contact is displayed on the Home and ...

Page 22: ...og interface setting The user can turn on or off the analog function separately Analog input parameters Control Range The actual output corresponding to the simulated quantity input generally 0 10V Return To Zero Voltage Current The return to zero voltage current of the analog input Note The analog input output function needs hardware support If the analog interface circuit is not configured at de...

Page 23: ...rotections Other Protects 1 Under Volt Under Curr Protection 2 Short Circuit Protect 3 Protection Switch Table 25 Other protects Other protections include under voltage protection under current protection short circuit protection and protection switches This type of protection is the opposite of the over voltage current protection mechanism that is when the output value is less than the protection...

Page 24: ...ction value Compared with the output value when the output value is less than the protection value the protection enters the pre triggered state Protection delay The time when the output is opened to effect the protection function Duration The time after the guard enters the pre triggered state to close the output and alarm fault occurrence Note 1 As the voltage current output behaves differently ...

Page 25: ... is 12345678 3 4 5 Reset 1 Factory Reset NO 2 Error Log Reset NO 3 System Data Reset NO 4 User Data Reset NO Table 29 Reset Recovery setting can reset clear a certain type of data Note Factory Reset reset data except running records and event records User Data Reset Reset some or all of the data such as communication Settings and function Settings to factory Settings ...

Page 26: ...21 4 Technical Specifications 4 1 Product parameters ...

Page 27: ...5 000A CAN RS485 Dry Node Analog OWP2040H 400 00V 12 000A CAN RS485 Dry Node Analog OWP2050H 500 00V 9 0000A CAN RS485 Dry Node Analog OWP2060H 600 00V 8 0000A CAN RS485 Dry Node Analog 3KW OWP3004H 45 000V 100 00A CAN RS485 Dry Node Analog OWP3006H 60 000V 80 000A CAN RS485 Dry Node Analog OWP3008H 80 000V 60 000A CAN RS485 Dry Node Analog OWP3010H 100 00V 45 000A CAN RS485 Dry Node Analog OWP301...

Page 28: ...S485 Dry Node Analog OWP8015H 150 00V 67 000A CAN RS485 Dry Node Analog OWP8020H 200 00V 50 000A CAN RS485 Dry Node Analog OWP8025H 250 00V 40 000A CAN RS485 Dry Node Analog OWP8030H 300 00V 34 000A CAN RS485 Dry Node Analog OWP8040H 400 00V 25 000A CAN RS485 Dry Node Analog OWP8050H 500 00V 20 000A CAN RS485 Dry Node Analog OWP8060H 600 00V 17 000A CAN RS485 Dry Node Analog OWP8100H 1000 0V 10 00...

Page 29: ...tting SAVE Save the normal data In normal mode Save App data In data mode RECALL Recall saved data on Home L R Switch local remote mode CONF Enter the function UI Figure 0 9 Enter figure Enter decimal point ENTER Enter the menu Input confirm Switch the home and the Auxiliary home Knob Clockwise Increment the input value numeric Settings Move Up N line Anticlockwise Decrease the input value numeric...

Page 30: ...me Use the clock or power on time as the reference time Time range Enable Enables or disables this time range On Off time set the time range Disable Parallel Series Connection Type of connection independent parallel or series Master slave the master or slave Independent Dry Contact output Control mode Disable local or remote control Relationship logic related to fault startup condition setting or ...

Page 31: ...ction detection delay 10 mS Time Of Duration Trigger Short circuit protection time 20 mS Protection Switchs Relevant protection switchs Password Password User Default Settings Reset Factory Reset Restoring factory Settings except for information records Error Log reset Clears fault Records System Data Reset Clears UI or all system setting User Data Reset Clears selected data 5 4 Warning list Name ...

Page 32: ... working process if it is start overshoot can set priority to voltage priority can also set Rise Time parameter to a reasonable value If the overshoot is steady state the over current protection value or overcurrent duration can be appropriately increased Over Curr P The software detects an over current error Confirm start overshoot or steady overshoot overshoot in working process if it is start o...

Page 33: ...tion in closed output state Operation after closing output RemoteCntr Comms Operate keys in remote mode Press L R to switch back to local control RemoteCntr Analog Switching Prior Cannot start output during priority switching Open output later Please Later The priority cannot be switched again during priority switching Wait 1 second and switch the priority again Step Mode Is En Cannot enable other...

Page 34: ... Set the power reference and start the output Illegal Data Saving a data group is invalid Save the data group after setting it correctly Full Data Space 128 data groups are full Delete redundant data groups and save them No Dada The precall data group is empty Call data after saving the corresponding data group AddrRange 1 247 Invalid MODBUS address Input valid address Func Code Err Communication ...

Reviews: