7.Appendix
Warning:
Disconnect the line cord at the rear panel and remove all test leads
connected to the instrument before replacing the line fuse. Failure
to do so could expose the operator to hazardous voltages that could
result in personal injury or death.
Use only the correct fuse type. Failure to do so could result in personal injury or
instrument damage.
Voltage
Fuse
90 - 240 V AC
250 V, F1AL
To perform the line fuse replacement, follow these steps:
1. Turn off the multimeter, remove all measurement leads and other cables from the
instrument, including the power cord.
2. Use a flat-blade screwdriver to remove the fuse box.
3. Replace the fuse with a new one, install it into the fuse box, and push the fuse box
back on to the rear panel.
41