Kymera 328i Quick Start Guide
Unpack the Boxes
Install Software (Solis or SDK2)
Kymera 328i
Spectrograph
1
You must have administrator access on your PC to perform this installation.
These instructions cover installation of both Andor Solis and Andor SDK if ordered. A copy of the
software is provided within the spectrograph box. Additionally, the latest software files are available
to download on request, please sign in or create an account at
andor.oxinst.com/downloads
Windows: Switch on the PC, run the setup exe from the supplied copy or the downloaded file: e.g.
AndorSDK2Setup-<version>.exe
1.
Confirm the version of software (minimum is v2.1.x).
2. Follow the on screen prompts.
3. Select the installation directory when prompted.
4.
If prompted for select the required hardware (spectrograph and/or camera) type.
5.
On the final window click on “Install”. Continue to follow the on screen prompts.
Kymera will require SDK2 or Solis installation, please see quick start guide for specific camera
model software requirements.
Online access to technical documentation is available at:
3
2
Camera(s)
(prefitted)
Grating Turret(s)
Slits/Coupling
Interfaces
Output Port
Adaptor(s)
Hex Key Set
Documentation
andor.oxinst.com/
welcome
Software
(copy supplied)
I
2
C Cable
USB 2.0 Cable
Power Supply
Please follow these guidelines when unpacking the Kymera:
1. The Kymera
should be placed in the environment it will be used in for a few hours before adjustment and first use.
2. Environment should be dust free.
3.
Spectrograph should be placed on a flat, stable support (e.g. optical table); Spectrograph feet are designed to be
secured on an optical table with appropriate screws.
The Kymera can be supplied with a range of accessories as specified at the time of ordering. Please check the
supplied items against the packing list to ensure all items are present.
Prepare Your Facility
Refer to the user manual for further information.
ATTENTION
Poids: >18kg.
Deux personnes sont
requises. Soulever avec
précaution - charge
inégalement répartie.
CAUTION
Weight: >18kg.
Two person lift required.
Take care when lifting -
unbalanced load.
Keep dry, <70% humidity
non-condensing.
Handle the spectrograph and
accessories with care.
Observe anti-static precautions
when installing the spectrograph,
camera and accessories.
Ensure sufficient clearance around
spectrograph and power supply
vents (Do
not
block vents when
operating on benchtop).
Operation of the spectrograph
close to intense pulsed sources
(e.g. plasma and X-ray sources)
may compromise performance.
Use correct power supply and cables.
Operate at a constant temperature
between 0°C and 30°C (temperature
fluctuation during operation may result
in calibration drift).
Store between -25°C and 50°C.
Please observe the following warning when
lifting the spectrograph: